• Keine Ergebnisse gefunden

Calibration-measurement unit for the automation of vector network analyzer measurements

N/A
N/A
Protected

Academic year: 2022

Aktie "Calibration-measurement unit for the automation of vector network analyzer measurements"

Copied!
4
0
0

Wird geladen.... (Jetzt Volltext ansehen)

Volltext

(1)

Adv. Radio Sci., 6, 27–30, 2008 www.adv-radio-sci.net/6/27/2008/

© Author(s) 2008. This work is distributed under the Creative Commons Attribution 3.0 License.

Advances in Radio Science

Calibration-measurement unit for the automation of vector network analyzer measurements

I. Rolfes1, B. Will2, and B. Schiek2

1Institut f¨ur Hochfrequenztechnik und Funksysteme, Leibniz Universit¨at Hannover, Appelstraße 9A, 30167 Hannover, Germany

2Institut f¨ur Hochfrequenztechnik, Ruhr-Universit¨at Bochum, Universit¨atsstraße 150, 44801 Bochum, Germany

Abstract. With the availability of multi-port vector net- work analyzers, the need for automated, calibrated measure- ment facilities increases. In this contribution, a calibration- measurement unit is presented which realizes a repeatable automated calibration of the measurement setup as well as a user-friendly measurement of the device under test (DUT). In difference to commercially available calibration units, which are connected to the ports of the vector network analyzer preceding a measurement and which are then removed so that the DUT can be connected, the presented calibration- measurement unit is permanently connected to the ports of the VNA for the calibration as well as for the measurement of the DUT. This helps to simplify the calibrated measurement of complex scattering parameters. Moreover, a full integra- tion of the calibration unit into the analyzer setup becomes possible. The calibration-measurement unit is based on a multiport switch setup of e.g. electromechanical relays. Un- der the assumption of symmetry of a switch, on the one hand the unit realizes the connection of calibration standards like one-port reflection standards and two-port through connec- tions between different ports and on the other hand it enables the connection of the DUT. The calibration-measurement unit is applicable for two-port VNAs as well as for multi- port VNAs. For the calibration of the unit, methods with completely known calibration standards like SOLT (short, open, load, through) as well as self-calibration procedures like TMR or TLR can be applied.

1 Introduction

For the automation of the calibration procedure of vector net- work analyzers, complex calibration units based on calibra- tion methods with completely known calibration standards, like SOLT (Short, Open, Load, Thru) are already commer-

Correspondence to: I. Rolfes (rolfes@hft.uni-hannover.de)

2 1

G1 G2 G3 G1 G2 G3

G1 G2 G3 G1 G2 G3

4 3

G1 G2 G3 G1 G2 G3

G1 G2 G3 G1 G2 G3 switchable 2-port-thru

Fig. 1. Example of a calibration unit for a 4-port-analyzer.

2 1

G1 G2 G3 G1 G2 G3

G1 G2 G3 G1 G2 G3

4 3

G1 G2 G3 G1 G2 G3

G1 G2 G3 G1 G2 G3 switchable 2-port-thru

DUT DUT

DUT DUT

Fig. 2. Setup of the calibration-measurement unit.

cially available (Henkel, 2006; Krekels and Schiek, 1995;

Krekels, 1996). In Fig. 1 a possible setup of a calibration unit for a 4-port-analyzer is shown. As can be deduced from the number of necessary calibration standards in Fig. 1, an automation of the calibration procedure is desirable and also recommendable in order to avoid calibration errors caused by e.g. connecting errors. The various standards are connected consecutively to the ports of the analyzer by a switch. For this purpose, the calibration standards have to be known ex- actly. After calibration the device under test (DUT) can be connected to the ports of the analyzer, and the scattering pa- rameters of the DUT can be determined error-corrected in relation to the chosen phase reference plane, e.g. in the phase plane of the DUT’s connectors.

Published by Copernicus Publications on behalf of the URSI Landesausschuss in der Bundesrepublik Deutschland e.V.

(2)

28 I. Rolfes et al.: Calibration-measurement unit

2 G

1 G

DUT DUT

Fig. 3. Combined calibration measurement setup based on a TLR- calibration.

Gk Hk

Gk Hk

Gk Hk

Gk Hk

Gk Hk

Gk Hk

l l

a b a’

G H

Fig. 4. Two-port error model of the VNA with TLR-calibration standards.

2 Calibration-measurement unit

For the automation of both the calibration of the analyzer and the measurement of the scattering parameters of a DUT, a combined setup, a so-called calibration-measurement unit is introduced. A setup for 4-port-measurements is shown in Fig. 2. The ports 1 to 4 are connected to the vector network analyzer and the DUT is connected to the ports named DUT.

Thus, a calibration-measurement unit results, where the cal- ibration standards do not have to be removed by the user in order to connect the DUT to the analyzer. The user can di- rectly connect the DUT to the resulting measurement ports, thus leading to a user-friendly setup, which can easily be re- calibrated without having to replace the DUT. However, as a precondition for this combined setup, the different switch po- sitions of a switch have to be symmetrical, so that the phase reference planes of the measured scattering parameters lie behind the switches in the phase plane of the DUT. As a con- sequence, the non-idealities of the switch are included in the error terms representing the systematic errors of the network analyzer. The calibration can either be performed based on calibration methods with completely known calibration stan- dards as e.g. SOLT as shown in Fig. 2 or with partly unknown calibration standards like TLR (Engen and Hoer, 1979; Eul and Schiek, 1991) as shown in Fig. 3 for a two-port setup. As the phase reference planes are preferably chosen in the plane

VNA

33

G 1 G

DUT

2

DUT

G DUT

GG

reference plane

Fig. 5. Extension of setup for multiport analyzers.

VNA

2 G

1 G

DUT DUT

reference plane

Fig. 6. Network analyzer with integrated calibration unit and mea- surement port for the contacting of the DUT.

of the DUT, for the TLR calibration a transformation of the phase reference plane is necessary.

This transformation can most easily be realized by choos- ing the reflection standard 0 for the TLR calibration as a short or open standard in the desired reference planesa,a0as illustrated in Fig. 4.

The error two-portsG andH in Fig. 4 are known after a TLR calibration. The error two-portsGk andHk for the transformed reference planesa,a0are calculable by

[Gk]= [G]

e−γ l 0 0 eγ l

(1) with

e2γ l= ρ

0 (2)

whereρ is the reflection coefficient of the TLR reflection standard in the reference planeb, which is known from self- calibration after the TLR calibration, and0is the reflection coefficient in reference planesa anda0, which here is cho- sen as0= −1 for the termination with a short. Based on Eq. (1) forGkand a similar equation forHk the phase refer- ence plane for the measurements can thus be moved into the phase plane of the DUT. The measured scattering parameters refer directly to the DUT without need for a deembedding.

Adv. Radio Sci., 6, 27–30, 2008 www.adv-radio-sci.net/6/27/2008/

(3)

I. Rolfes et al.: Calibration-measurement unit 29

Hk Bk

Hk B

Hk B

Hk Bk

Hk Bk

lV

lV

C C’

Fig. 7. Two-port setup for elimination of influence of additional cables.

Fig. 8. Photo of the calibration measurement unit.

%vspace-3mm

The TLR-calibration setup of Fig. 4 as well as the previ- ously presented calibration setups based on SOLT can eas- ily be adopted to multiport analyzer systems by adding fur- ther line connections to the calibration measurement unit as shown in Fig. 5. In case that these lines are of unknown length but of identical geometry and line impedance as the previously used line standards for TLR calibration, it is pos- sible to determine this unknown line length from an addi- tional reflection measurement.

3 Integration into an analyzer system

The calibration standards can be integrated completely into a network analyzer system, so that only the measurement port for the contacting of the DUT remains as an interface to the user. This is demonstrated in Fig. 6. This setup allows an au- tomated calibration of the measurement system without be- ing in need for a manuel interaction of the user. On the one hand, this helps to reduce the required knowledge about cal- ibration technology of the user and on the other hand, this possibility of automated calibration and combined measure- ment can furthermore be useful for applications where no user interaction is possible in the measurement plane, as e.g.

in industrial drill hole applications where the measurement plane can be deep under earth’s surface.

0 2 4 6 8 10 12 14 16 18

-50 -40 -30 -20 -10

S11 0

dB

f / GHz S11

dB

Fig. 9. Verification of the symmetry of the switch. Comparison of S11measurements for all 6 positions of the switch.

f / GHz

0 2 4 6 8 10 12 14 16 18

-0.8 -0.6 -0.4 -0.2 0

S21 0.2

dB

f / GHz

0 2 4 6 8 10 12 14 16 18

-0.8 -0.6 -0.4 -0.2 0

S21 0.2

dB S21

dB

Fig. 10. Verification of the symmetry of the switch. Comparison of S21measurements for all 6 positions of the switch.

For the case that additional cables are necessary for the contacting of the DUT, it is possible to eliminate the influ- ence of the cables by performing two additional measure- ments: first, the measurement of the transmission of the ca- bles and secondly, the measurement of the cables with a re- flection standard placed in the phase reference plane. The setup is shown exemplarily in Fig. 7. Similarly to the pre- viously described multiport extension, the influence of the cables can be corrected.

4 Measurement results

In order to verify the functionality of the proposed calibration-measurement unit, a demonstrator was realized on the basis of electromechanical relays as shown in Fig. 8.

For verification of the symmetry of a switch, the different po- sitions of a switch are compared. The measurements of the scattering parametersS11 andS21in Figs. 9 and 10 show a good agreement of the high-frequency behavior for the dif- ferent switching positions.

In Figs. 11 and 12, some further results for the verification of the switches symmetry are given. In this case, the scat- tering parameters are measured for varying terminations like open, short, match and the repetition of contacting.

In Figs. 13 and 14, the calibrated measurements of the scattering parameters S11 and S21 of a 3dB-attenuator are

www.adv-radio-sci.net/6/27/2008/ Adv. Radio Sci., 6, 27–30, 2008

(4)

30 I. Rolfes et al.: Calibration-measurement unit

0 2 4 6 8 10 12 14 16 18

-50 -40 -30 -20 -10

S11 0

dB S11

dB

f / GHz

Fig. 11. Verification of the symmetry of the switch. Comparison of measurements ofS11while contacting varying terminations like open, match and repetition of contacting.

f / GHz

0 2 4 6 8 10 12 14 16 18

-0.8 -0.6 -0.4 -0.2 0

S21 0.2

dB S21 dB

Fig. 12. Verification of the symmetry of the switch. Comparison of measurements ofS21while contacting varying terminations like open, match, and repetition of contacting.

shown exemplarily. A comparison of a direct measure- ment of the DUT and the measurement with the help of the calibration-measurement unit proves the robust functionality of the developed method.

5 Summary

A calibration-measurement unit is presented which enables on the one hand automated, repeatable calibrations of vec- tor network analyzers and on the other hand the measure- ment of scattering parameters of DUTs. Well-known calibra- tion methods like SOLT as well as self-calibration methods like TLR are applicable. The possibility of integration of the whole setup is described, leading to a measurement system where the user can directly connect the DUT to the measure- ment ports of the system, while the calibration standards can be left in the system so that a repetition of calibration is possi- ble at every time. The setup is based on switching multiports, which have to be symmetrical individually for the proposed setup. For the verification of the robust functionality of the calibration-measurement unit a demonstrator and measure- ment results are presented.

S11 dB S11 dB

0 2 4 6 8 10 12 14 16 18

-50 -40 -30 -20 -10 0

f / GHz

Fig. 13. Comparison of the measurement ofS11of a 3dB-attenuator based on a direct TLR-calibration (blue line) and based on the calibration-measurement-setup (red line).

0 2 4 6 8 10 12 14 16 18

-4 -3.8 -3.6 -3.4 -3.2 -3 -2.8 -2.6 -2.4 -2.2 -2

f / GHz S21

dB S21 dB

Fig. 14. Comparison of the measurement ofS21of a 3dB-attenuator based on a direct TLR-calibration (blue line) and based on the calibration-measurement-setup (red line).

References

Engen, G. F. and Hoer, C. A.: Thru-Reflect-Line: An improved technique for calibrating the dual six port automatic network an- alyzer, IEEE T. Microw. Theory, 27, 987–993, 1979.

Eul, H. J. and Schiek, B.: A Generalized Theory and New Calibra- tion Procedures for Network Analyzer Self-Calibration, IEEE T.

Microw. Theory, 39, 724–731, 1991.

Henkel, A.: Unrivaled – up to eight test ports in a single unit, News from Rohde & Schwarz, 189, 26–29, 2006.

Krekels, H.-G.: AutoKal: Automatic Calibration of Vector Network Analyzer ZVR, Rohde & Schwarz Application Note 1EZ30 IE, 1996.

Krekels, H.-G. and Schiek, B.: A novel Procedure for an auto- matic Network-Analyzer Calibration, IEEE T. Instrum. Meas., 44, 291–294, 1995.

Adv. Radio Sci., 6, 27–30, 2008 www.adv-radio-sci.net/6/27/2008/

Referenzen

ÄHNLICHE DOKUMENTE

We have developed a Bayesian framework to be applied for the analysis of the LPF mission planned system identification experiments.. Different approximations have been employed

5 shows the comparison of experimental sensor data of Rock 1 during the RF04 cam- paign with simulation output using the best parameter fit of the previously conducted parame-

As predictable, from the flume tests it was derived that the run-out behavior is controlled by both kaolin and sand concentrations, but here, for lack of space, we comment only

Case 2 The second example is a set of bi-component agglom- erates examined in Dosta et al. In this study, the analyzed materials consist of one type of primary particles and two

width sequence (dashed magenta); the observations (grey, lower left panel), the adaptive estimation of 0.75 quantile (dotted black), the true curve (solid black), the quantile

Over the horizon- tal formula (3.23), now the height resulting from it can be calculated less the real height (MSL) as ∆ h. This height is positioned in relation to the

This calibration certificate documents the traceability to national standards maintained by the DAkkS or other national standards, which realize the physical units

The purpose of the present work was the evaluation of the performance characteristics of various casein cali- bration models for FT-IR measurements using sets of commercial