• Keine Ergebnisse gefunden

4.3 Defocused Imaging

4.3.1 Theory

We start our theoretical considerations from the derivation of the transmitted electric field of a dipole on a dielectric interface from section 2.4.2. In particular, we refer to the Weyl representation of the transmitted field (2.131)

ET(r) = ik02

¨ dq w1

ˆe+2pTp(ˆe+1p·p) +ˆesTs(ˆes·p)

ei[q·(en−en0)+w1|z0|+w2z]. (4.15) Here we use the same sense of orientation and conventions as previously used: The positive direction of z is downwards into the medium n2, z > 0 below the interface.

The dipole is present in medium n1 at a position (˜n0, z0), z0 < 0. Since only the far-field emission plays a role here, we consider the components with |q| ≤ n1k1 only, neglecting all imaginary solutions. The electric field for a vertically oriented dipole (p =pˆz) contains only p-waves. The area element dq can be treated as dq=qdqdψ = k2w2sinθ22dψ. Therefore, the magnitude of the electric field at an emission angle θ2 per solid angle dΩ = sinθ22dψ is given by

Ep2) =ˆe+2pEp2) = −ˆe+2pqw2n2

w1n1 Tpeiw1|z0|. (4.16)

CHAPTER 4. SM ORIENTATION 4.3. DEFOCUSED IMAGING

Figure 4.26: A schematic showing the geometry of vectors considered in this section. The dipole is located at a position (ρ0, z0) with an orientation (β, α). The interface separates the two media n1 =

1 andn2 =

2 as shown. The unit vectorsˆe+ip represent the directions ofp-waves, and ˆes

fors-waves. θ1 andθ2 are the angles the vectors k+1 and k+2 make with respect to the normal of the interface, andψ is the angle between the plane of incidence and thex-axis on the interface as shown.

For the case of a parallel dipole, one has bothp- ands-waves in the transmitted electric field. The field not only depends on the angle with respect to the optical axis θ2, but also changes with the azimuthal angle ψ. We therefore, split up the field into two components

ˆ

e+2pEpk2) cosψ =ˆe+2pw2n2

w1 Tpeiw1|z0|cosψ, (4.17) and

ˆ

esEsk2) sinψ =−ˆesw2n2

w1 Tseiw1|z0|sinψ. (4.18) Thus, for every dipole oriented at an angle (β, α) whereβis the angle from the positive x-axis andα, the inclination with respect to the optical axis, the magnitude of the electric field at direction (θ2, ψ) per solid angle can be written in terms of the components of a vertical and parallel dipoles.

E(θ2, ψ) = ˆ

e+2pEpk2) cos(ψ −β) +ˆesEsk2) sin(ψ−β)

sinα+ˆe+2pEp2) cosα (4.19) Let us now consider the imaging optics of the setup. A geometrical depiction is provided in figure 4.26. A defocusing can be achieved in two ways: i) By shifting the

4.3. DEFOCUSED IMAGING CHAPTER 4. SM ORIENTATION

Figure 4.27: The geometry of the imaging setup. The dipole is placed on the optical axis in a medium of refractive indexn1 at a distancez0 from the interface atz= 0. The medium between the interface and the focal sphere in the object space has a refractive indexn2. The focal sphere in image space and the image plane are present in a medium of refractive index n0. R is the vector from the focal point of the objective to a point on the focal sphere in object space,|R| is the focal length of the objective.

It is marked here as the radius of the focal sphere in the object space with center at the point of the intersection of the interface with the optical axis. Similarly, |R0|represents the focal length in image space.

objective, and therefore its focal plane, relative to the interface. Remember here that the molecule under consideration is present at a distance |z0|from this interface. ii) By placing the detector away from the focal plane. In a wide-field fluorescence microscope such as that considered in chapter 3, these two ways produce the same result and are completely identical. Here, in this section, we will specifically consider the case where the detector is shifted from its imaging plane. The mathematics of the image formation remains, once again, completely identical to the work of B¨ohmer et al. [59]. As we saw in section 4.1, the electric field around the optical axis in the image plane, considering aberration free optics, is described by the integral formula from Richards and Wolf [131]

E(r0) = −i|k2| 2π

¨

a(s0x, s0y)eik0s0·r0)dΩ0 (4.20) where the solid angle element is dΩ0 = ds0xs0ds0y

z , and the integration extends over the complete angular space Ω= (θ20, ψ); ψ varies from 0 to 2π and θ02 from 0 to θmax0 . The relationship between θ2 and θ20 is given by Abbe’s sine condition, n2sinθ2 =M n0sinθ20; and therefore, θ0max = arcsin(N.A./M n0), where N.A. denotes the numerical aperture of the objective and n0 is the refractive index of the imaging medium, usually air, and M is the magnification of the system. k2 and k0 are the wave vectors of light in the object (in glass) and image space, respectively. Vector a represents the strength vector right at the ‘focal sphere’, i.e. the sphere with center on the imaging plane at the optical axis (see figure). r0 is the coordinate of the target point in the image plane from the focal sphere, where the electric field is being calculated. Thus, for a vertical dipole, it is given by equation (4.16) and for a parallel dipole by the sum of both p- and s-fields

CHAPTER 4. SM ORIENTATION 4.3. DEFOCUSED IMAGING

presented in equations (4.17) and (4.18). The main idea behind equation (4.20) is to present the electric field magnitude in the image space as a superposition of plane waves.

The field magnitude for a vertical dipole at a plane before or after the focal plane can be expressed, by slightly modifying this equation as

E(r0) = M target point on the plane in cylindrical coordinates centered at the point of intersection of the optical axis with the focal plane in image space. Therefore,ρ0 is a vector pointing to the target point from the optical axis, z0 is the distance of the target point from the focus along the optical axis andφ0 is the angle betweenρ0 andˆex. The square root factor in equation (4.22) together with magnificationM ensure energy conservation [131]. The connection between the coordinates of the position of the dipole in object space and image space are found using the relations ρ0 = Mρ, for the distance away from the optical axis and, z0 = M2z along the optical axis. These relations also state that in order to achieve a defocusing ofδz in object space, one must displace the detector by a distanceM2δz in the image space and vice-versa.

Similar relations for the amplitude of the magnetic field in the image space can be found:

4.3. DEFOCUSED IMAGING CHAPTER 4. SM ORIENTATION

The integration over ψ can be performed analytically. The electric field magnitude for a dipole oriented at (β, α), writing the electric and magnetic fields as a combination of parallel and vertical dipoles (see equation (4.19)), is given by

(Ej Jnare Bessel functions of the first kind ofn-th order, with functional argumentk0ρ0sinθ200 =|ρ0|). The final position dependent intensity in the image plane is given by thez component of the Poynting vector

S = c

8πˆez·Re{E×B}. (4.29)

Note that these equations are written considering a two lens system with magnification M. The derivation for a four lens system remains the same except for the fact that the images will be inverted in the xy-plane. Therefore, one must then replace φ0 with −φ0 in the final equations for ej and bj. Intensity patterns for a few orientations of the pem for a four lens system are calculated and shown in figure 4.28.

CHAPTER 4. SM ORIENTATION 4.3. DEFOCUSED IMAGING

Figure 4.28: Calculated intensity images of a dipole on a camera plane with a defocusing value of δz= 0.9µm in the object space. The emission wavelength was set toλem = 690 nm The N.A. of the objective for the calculations was set to 1.49, with a net magnification ofM = 200. The refractive index in image space was set ton0 = 1. The camera pixel size was 13µm×13µm. The dipoles were assumed to be on top of an air/glass interface (z0= 0) and on the optical axis (n1= 1;n2= 1.52). Each pattern spreads over 20×20 pixels on the detector. Note that the images show normalized intensities. From top left to bottom right,αandβ values for the calculated dipole patterns are:

Column/Row 1 2 3 4 5

α, β α, β α, β α, β α, β

1 90,0 90,45 90,90 90,135 70,0 2 70, 45 70,90 70, 135 70,180 70,225 3 70,270 70,315 50,0 50,60 50,120 4 50,180 50,240 50, 300 30,0 30,90

5 30,180 30,270 10,0 0,0 −,

4.3.2 Experimental Setup

A wide-field based imaging setup has been described in detail in section 3.1.2 earlier. A total illumination area of 80µm×80µm in the object space, an exposure time of∼ 3 s and an average illumination power of 0.2 kWcm−2 gave excellent signal-to-noise ratio in the defocused intensity patterns for single molecules (see figure 3.8). In this section, we will describe mainly the setup for measuring the defocused patterns using a focused illu-mination and a camera that has been displaced from the focal plane in the image space.

4.3. DEFOCUSED IMAGING CHAPTER 4. SM ORIENTATION

The excitation system and sample preparations were described in section 4.1.2. The fluorescence collected by the objective was recollimated using an achromatic doublet (AC254-150-A, Thorlabs Inc.) and any backscattered laser was blocked using a long-pass filter (EdgeBasic BLP01-647R, Semrock). The pinhole, between the tube lens and the achromatic doublet, was omitted for the sake of alignment simplicity. In principle, a pinhole with a diameter of 100µm to 150µm diameter should not alter the imag-ing properties of the setup at all, except in contributimag-ing to the blockimag-ing of unwanted background arising due to the presence of molecules above and below the excitation focus if any. In the limit that the emitters are well scattered in space and separated from each other, the pinhole does not make any significant difference and just adds to the alignment efforts. After recollimation and filtering, the emission light was reflected by a custom designed flipping mirror assembly into the defocused imaging detection system (see figure 4.7 for a complete design of the setup). An additional band-pass filter was inserted in the beam path in order to narrow down the spectral range of the detected photons (FF01-692/40 Semrock, for Atto 655 molecules). Thereafter, the light was focused on an EMCCD camera (iXon DU860-D, Andor) using a large focal length lens (AC508-1000-A, Thorlabs Inc.) mounted on an xy-translation mount (LM2XY, Thorlabs Inc.). The net magnification of the setup was M = 400 which, taking into account the pixel size of the camera (24µm×24µm), results in an area of 60 nm×60 nm per pixel in the object space. In order to achieve a defocused image plane, the camera was mounted on a stage which was placed in between two rails, facilitating the change of position by sliding (see figure 4.29). For a defocusing of −0.9µm in object space (δz = 0.9µm), the camera was shifted by a distance of 14.5 cm (M2δz) towards the lens from the imaging focal plane. The focal plane in image space was located by focusing the laser on the air/glass interface of an empty coverslip, imaging the back-reflected light onto the camera chip without any filters (or with a neutral-density filter if the back-reflected intensity was too high), and sliding the camera along the rails such that the image shows a Gaussian spot with the narrowest width (Note that due to the high magnification and the M2 relationship between the axial distances in the image space and object space, an error of 1 mm translates to ±6.25 nm in object space, which is negligible compared to the size of the excitation volume). Depending on the brightness and stability of the molecules, the laser power and exposure time were adjusted. For example an excitation power of 0.5 kWcm−2 was used together with an exposure of 5 s for Atto 655 molecules at glass/air interface.