Magnetic performance measurement

A magnetic performance measurement was carried out, in order to get values of the real screening performance of the septum YIN (23.12.1999). Therefore the septum pulser was connected to the magnet and the vacuum tank was vented with liquid nitrogen. Then the septum exit blind flange could be opened under nitrogen atmosphere. A translucent plastic bag, connected to a membrane pump and equipped with rubber gloves was used therefore.

Fig. 1: bag set-up (w/o magnetic field probe and pump)

With a specially prepared miniature magnetic field probe the main field and the leakage field have been measured subsequently, along the magnetic pole of the septum. It was found that the magnetic screening at the ends of the septum exceeded 1/1000. Over a distance of about 10mm the magnetic field increased up to 3/1000 of the main magnetic field in the center of the septum magnet.

(2)

Fig. 2: Additional CO-Netic screen at the septum entrance

Fig. 3: Two additional magnetic screens at the septum exit

Then custom made additional CO-Netic screens were added. Two screen pieces at the exit of the septum and one at the entrance improved the screening as is shown in Fig 4.

(3)

Fig. 4: Leakage Field with and without CO-Netic screen

Referenzen

ÄHNLICHE DOKUMENTE

Auf der Basis eines breit herausgearbeiteten Grundverständnisses von Performance Measurement entwickelt Robert Hauber allgemeingültig die Konzeption eines Perfor- mance

6.6.3 Entwicklung eines angepassten Performance Measurement- Frameworks für

In terms of the strategic innovation management literature the discussion section has shown that the findings are compliant with the conceptual insight from the strategy of superior

wie sich mit Hilfe des CRM Performan- ce Model (CPM) die zentralen Erfolgs- größen des CRM unter Berücksichtigung ihrer unterschiedlichen Wirkdimensio- nen messen lassen und somit

In summary, the initial method evaluation indicates that both the MetricML language concepts and the corresponding procedural guidelines of MetricM promise to appropriately support

The test conditions required for Durability and Drop- Out Activity evaluation are quite different. In the case of durability, the selection of the tape movement and the

In addition to our measurements of the magnetic moments of a large number of bacteria in various external magnetic fields, we also determined the saturation magnetic moment of

The following table provides an overview of the individual RCI change scores at the points in time when measuring was carried out (henceforth, ‘time of measurement’ [T]) as indicated