Dr. Torben Gentz

Postdoc, Marine Geochemistry Alfred Wegener Institute

Buenos Aires; November 10, 2014

High resolution detection of the dissolved gas composition in

aquatic systems by novel underwater mass spectrometry.

(2)

Water column sampling

Phase separation:

gas phase from aqueous phase

Headspace technique for analysis of discrete

samples

Gas analysis by gas chromatography

Problems:

-time consuming

-coarse spatial and temporal resolution

(Lammers and Suess 1994)

E

STABLISHED METHODS FOR HYDROCARBONS

(3)

U

NDERWATER

M

ASS

S

PECTROMETER

(I

NSPECTR

200-200)

Mechanical

• Dimension:

• UWMS housing: ø19.5 cm x 120 cm

• Cryotrap housing: ø19.5 cm x 30cm

• Weight Air - 38.5kg

Water – neutrally buoyant

• Max .operating depth: 200m standard housing

• Construcion hard anodized aluminium

Electrical

• 24 volts DC external power

• 95 Watts power consumption

• Embedded Pentium™ PC controller

• LAN communications port Mass Analyzer

• Quadrupole based

• Membrane introduction source

• Vacuum: Varian turbo molecular with diaphragm backing pumps

SPECIFICATIONS

(4)

• Both methods are comparable

• No sampling preparation

• Simultaneous measurement of the dissolved gases

• No artefacts during sampling

• Up to 750 times higher sampling frequency

Higher temporal and spatial resolution

Inspectr200-200 vs. GC

C

OMPARISON OF THE

I

NSPECTR

200-200

VS

.

CONVENTIONAL TECHNIQUES

(5)

E

XAMPLE OF

G

AS MEASUREMENTS BY

UWMS

(6)

M

ODE OF OPERATION

Laboratory measurements Ex situ In situ in a frame

including benthic chamber

In situ at sediment- water-transition-zone

(7)

over 250.000 samplings in 3 years

R

ESULTS

: Ex situ by pumping on board

„Grundfos“ pump

• Up to 80 m water depth

Methane concentration in Lake Constance

(8)

11900 samples in various depth in between 24 hours

R

ESULTS

: In situ in a frame

UWMS in a frame

• Online up to 100 m water depth

• Offline up to 200 m water depth

• In situ benthic chamber measurements

• Cruise vessel needed

Methane concentration above a pockmark in the North Sea

(9)

UWMS on top of a human drawn vehicle and at quay

• Battery or external power supply

• Measurements at the seafloor

• In situ benthic chamber measurements

• Only <2 m water depth

R

ESULTS

: In situ at the sediment-water-transition-zone

UWMS on top of the MarBot

Methane concentration in the Baltic Sea

(10)

R

ESULTS

: Laboratory measurements

Sediment core

Benthic chamber measurements with the UWMS and GC in the Baltic Sea

Laboratory analysis

• No power restriction

days of operation time

• Simultaneous observation of the dissolved gases

(11)

Thanks for your attention!

Questions?

Referenzen

ÄHNLICHE DOKUMENTE

H 2 SOM: Hierarchical hyperbolic self-organizing map; H&amp;E: Hematoxylin and eosin; GRINE: Analysis of GRaph mapped Image data NEtworks; LDA: Latent dirichlet allocation; m/z: Mass

The fractions of hexane and dichloromethane extraction from marupa´ (Simaruba amara) and (Bertholletia excelsa) leaves were analyzed by HT-HRGC (high temperature high resolu- tion

The experimental results are in perfect agreement with the matrix IR spectra, mw measurements and ab initio calculations but disagree with a recently claimed low resolution detection

Alberto dos Santos Pereira a , Moˆnica Freiman de Souza Ramos b , Elisa Suzana Carneiro Poc¸as a , Patricia Castro Moreira Dias b , Elisabete Pereira dos Santos b , Joaquim

Therefore, with respect to real estate similar conflicts can be observed in South Bavaria to the types of conflict discussed in the empirical studies on other Western

The optical baseline design meets all the requirements imposed by the Crires spectrograph (Sec. 11.2), and delivers diffraction limited image quality at the entrance slit. While the

The only value that is used for the control process is the depth x 1 , which is also available on the real system trough the pressure sensor.. The triangular blocks represent

All components of the 20S proteasome of mouse intestine separated by 2-D gel electrophoresis were unambiguously identi- fied by UV-MALDI FTICR-MS following tryptic in-gel