Robotic devices for paediatric rehabilitation:

a review of design features

Alberto Gonzalez

1

, Lorenzo Garcia

1*

, Jeff Kilby

1

and Peter McNair

2

Abstract

Children with physical disabilities often have limited performance in daily activities, hindering their physical development, social development and mental health. There- fore, rehabilitation is essential to mitigate the adverse effects of the different causes of physical disabilities and improve independence and quality of life. In the last decade, robotic rehabilitation has shown the potential to augment traditional physical reha- bilitation. However, to date, most robotic rehabilitation devices are designed for adult patients who differ in their needs compared to paediatric patients, limiting the devices’

potential because the paediatric patients’ needs are not adequately considered. With this in mind, the current work reviews the existing literature on robotic rehabilitation for children with physical disabilities, intending to summarise how the rehabilitation robots could fulfil children’s needs and inspire researchers to develop new devices.

A literature search was conducted utilising the Web of Science, PubMed and Scopus databases. Based on the inclusion–exclusion criteria, 206 publications were included, and 58 robotic devices used by children with a physical disability were identified.

Different design factors and the treated conditions using robotic technology were compared. Through the analyses, it was identified that weight, safety, operability and motivation were crucial factors to the successful design of devices for children. The majority of the current devices were used for lower limb rehabilitation. Neurological disorders, in particular cerebral palsy, were the most common conditions for which devices were designed. By far, the most common actuator was the electric motor. Usu- ally, the devices present more than one training strategy being the assistive strategy the most used. The admittance/impedance method is the most popular to interface the robot with the children. Currently, there is a trend on developing exoskeletons, as they can assist children with daily life activities outside of the rehabilitation setting, propitiating a wider adoption of the technology. With this shift in focus, it appears likely that new technologies to actuate the system (e.g. serial elastic actuators) and to detect the intention (e.g. physiological signals) of children as they go about their daily activi- ties will be required.

Keywords: Robotic, Exoskeletons, Rehabilitation, Assistance, Children, Physical disability

Open Access

© The Author(s), 2021. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the mate- rial. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://

creat iveco mmons. org/ licen ses/ by/4. 0/. The Creative Commons Public Domain Dedication waiver (http:// creat iveco mmons. org/ publi cdoma in/ zero/1. 0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

REVIEW

*Correspondence:

lorenzo.garcia@aut.ac.nz

1 BioDesign Lab, School of Engineering, Computer and Mathematical Sciences, Auckland University of Technology, Auckland, New Zealand

Full list of author information is available at the end of the article

(2)

Introduction

Mobility and exploration are essential in children’s development and contribute towards cognitive, physical, social and emotional development. However, children with physical disabilities present limitations when performing activities autonomously, which hinders their typical development [1]. Ongoing paediatric physical disability arose from many different causes, including neurological disorders like cerebral palsy (CP) [2], Stroke [3]

and acquired brain injury (ABI) [4], neuromuscular diseases such as Duchenne muscular dystrophy (DMD) [5] and spinal muscular atrophy (SMA) [6], or traumatic injuries [7, 8]

(Table 1).

Rehabilitation is essential to help the children recover or maintain functionality when interacting with their environment, improving the quality of life and autonomy [9, 10].

Furthermore, early access to rehabilitation is critical for children while they are in the stage of development. The gait pattern and motor abilities are still malleable [11], intend- ing to reduce the probability of developing more severe levels of disability [12, 13].

The standard therapies to manage the musculoskeletal system’s deterioration and improve and maintain physical ability include passive orthoses, surgery, and physio- therapy [15, 16]. Physiotherapists prescribe, monitor, and guide exercise, which can pre- vent an unnecessarily sedentary or immobile lifestyle. The most extensively investigated aspect of physiotherapy is the effect of direct interventions on upper or lower limbs.

Such interventions often involve intensive stretching and strengthening exercises facili- tated by the physiotherapist [2] to improve motor skills. These interventions are often highly labour intensive and can be challenging to perform [17]. Furthermore, the effec- tiveness of physiotherapy often depends on the experience of the physiotherapist. Thus, it is not easy to achieve optimal consistency and repeatability between rehabilitation ses- sions [18, 19].

There is a growing interest in robots that can support the patient, the family and the medical professional in a wide range of activities used for the care of people with physi- cal disabilities, for example, companion robots [20, 21], monitoring robots [22] and sur- gery robots [23], all of them can be considered as healthcare robots.

Healthcare robots can be divided into three main categories, clinical robots, assistive robots and rehabilitation robots [24, 25]. Clinical robots are focus on supportive care and cure process (e.g. help in surgery and diagnosis) in clinical environments; assis- tive robots primary function is to provide assistive help either to carers or directly to patients either in a hospital or in a specialist care facility (e.g. patient lifting and to assist in routine services); rehabilitation robots are robots design towards restoring the func- tionality and mobility of people with physical disabilities, in that case, the recovery of

Table 1 Incidence or prevalence of conditions that cause physical disabilities in children

Condition Incidence or prevalence

Cerebral palsy Prevalence of 1 per 500 live births [2]

Stroke Incidence of 1.2 to 13 per 100,000 children per year [3]

Traumatic brain injury Incidence of 691 per 100,000 children [7]

Duchenne muscular dystrophy Prevalence of 1 per 5000 live male births [14]

Spinal muscular atrophy Prevalence of 7.8–10 per 100,000 live births [7]

(3)

mobility could be achieved by assisting the patient during ADLs (e.g. walking and grasp- ing objects) [17, 26] or with physical training therapy [24, 27–30], and are the main focus of this study.

Rehabilitation therapy for the recovery of mobility based on robots has been proposed as a new procedure for children with physical disabilities [31]. This robot-assisted reha- bilitation therapy consists of a mechatronic device that provides highly repetitive and task-specific guided movements autonomously [32, 33]. The use of robots in rehabilita- tion therapies bring advantages over traditional therapies, as they allow extensive prac- tice in children with substantial disabilities, reduced effort required of therapists during the exercises, and provide a quantitative assessment of the patient’s motor function (e.g.

quantitative feedback of range of motion (ROM) and strength with each repetition) [34–38].

Rehabilitation robots are often classified by their mechanical structure and are gener- ally divided into end-effectors and exoskeletons [28, 39]. End-effector devices work by applying forces to the distal segments of limbs, creating what is termed a “mechanical chain” that prompts movements of other parts of the limb generating a pattern of spe- cific activity across different joints. If utilised on a single segment and joint, their simple structure makes it easier to adapt them to many patients and needs less complicated control algorithms. However, it is difficult to isolate specific joints since they produce complex movements that involve the whole limb [40].

Contrarily, robotic exoskeletons could be termed “wearable machines” that mirror the patient’s skeletal structure; therefore, they only move the joint of the limb where the exoskeleton is worn. This approach allows for independent and concurrent control of specific segments of the limb. However, it is essential to adjust the length of sections of the robot to the lengths of the segments of the patient limb. Moreover, when the joint is in motion, the position of the centre of rotation can change, creating discomfort in the user. Thus, increasing the number of degrees of freedom of the robot increases the con- trol algorithm’s complexity, weight, mechanical complexity, and power requirements, making it unattainable for home use [18, 41].

Apart from mechanical structure, robots possess essential elements to ensure the sys- tems’ reliability and robustness [42]. Actuators, training strategy and the Human–com- puter interface (HCI) are among these essential elements. The actuators play a crucial role because they determine the torque and movement provided by the robot and influ- ence the total weight and compliance of the system [29, 43]. The training strategy and the HCI are an integral part of the robot-assisted rehabilitation since it determines how the patient interacts with the robot and the type of assistance that the robot can pro- vide. Many authors have analysed these last two characteristics as part of the robots con- trol [28, 30, 44]. However, control also involves “low level” considerations that are more related to the internal communication of the components (sensors, structures, micro- controllers, actuators, etc.) at a hardware level rather than how the device interacts with the patient [45–49].

Although multiple devices for the robotic rehabilitation of upper and lower limbs have

been developed, at least in a proof-of-concept phase [24, 39, 42, 44, 45, 50, 51], most

presented robots were designed for adult users, impeding their use on the paediatric

(4)

population. For example, commercial exoskeletons are made for a subject 150 cm tall onwards [52], while the average height for a 5-year-old child will be around 110 cm [53].

However, to develop technology planned to be used on the paediatric population is not only a matter of reducing the size of the robots. But it should be tailored to their own capabilities and goals that differ from those of the adults. For instance, a simple downs- caling of the robots is not enough as the normalised joint torques on adults are greater than those of a child [54, 55], making them potentially dangerous when used on small children. Additionally, in the case of children, as their cognitive abilities are still devel- oping, it could be hard for them to fully understand how the technology works [52, 56].

Hence, it is hard to adapt a robot made for adults to be used by children since the robots do not fulfil the children’s needs [26, 57].

Consequently, to address the children’s needs adequately, it is essential to include them and other stakeholders (e.g. family members, clinicians, and health care providers) dur- ing the development process, providing feedback to identify possible issues of impor- tance [1, 43]. Furthermore, it is essential to focus not only on addressing the impairment or limitation in users’ functional abilities, but also on other fundamental needs, like accessibility and aesthetics [58], to avoid the user abandoning the rehabilitation device due to frustration [59].

Despite the progressive development of robotic rehabilitation devices, their applica- tion to the paediatric population is still scarce. Consequently, the key features to design an optimal robotic rehabilitation device that better enhance children’s abilities with physical disabilities have not been well defined yet. Based on this framework, this review aims to address the following questions: (1) What are the design requirements for paedi- atric rehabilitation robots? (2) How does the current technology contribute to achieving the paediatric design requirements? And (3) How do the paediatric conditions impact the device design?

Methods

An in-depth literature search was performed to conduct the review, following the search strategy of the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines [60].

A literature search was conducted to identify literature associated with the topic based on searches in PubMed, Scopus, and Web of Science, using the combination of the fol- lowing keywords: (pediatric OR kid* OR child*) AND (aid OR assist* OR improve* OR augment* OR enhance* OR reinforce* OR therap* OR rehabilitation) AND (active ortho*

OR exoskeleton* OR wearable robot* OR portable robot* OR robot* suit OR robot*) AND (movement OR motion OR walk* OR gait OR grasp* OR handl*). To make our search as complete as possible, a search through the university library databases was also conducted.

After the preliminary search, the following inclusion and exclusion criteria to narrow the literature search were used. The inclusion criteria were:

1) Studies involving robotic devices for robot-assisted rehabilitation therapy,

2) Studies involving robotic devices for assessment of patients with Physically disabili-

ties,

(5)

3) Studies involving devices designed for children or utilised with a paediatric popula- tion (< 18 years old),

4) written in English, 5) full-text articles.

And the exclusion criteria were:

1) Studies that only present software solutions or simulations, 2) Studies involving passive devices (do not have actuators), 3) Studies involving postural change,

4) Studies involving only the adult population and

5) Studies involving robots that do not replace the movement itself (e.g. wheelchairs).

Findings

The outcome of this literature review is compiled in the following sections:

• An overview of the literature search,

• the paediatric robotic rehabilitation design requirements,

• an analysis of the type of robots used in paediatric robotic rehabilitation;

• the actuators to drive the robots;

• training strategy of the robots;

• the human–computer interface of the assistive systems, and

• the treated conditions in children with physical disabilities.

Literature search

Based on the keywords mentioned in the methods section, 1604 publications were found, with:

• 811 publications from Web of Science,

• 547 publications from PubMed,

• 241 from Scopus, and

• 5 from a search on the University library.

First, a check was made for duplicated publications. After this process, the abstracts of 1248 publications were screened, and 301 titles were selected for full-text reading. After carefully applying the inclusion–exclusion criteria to the full read papers, 206 publica- tions were selected. Among the chosen publications, 10 were reviews, 42 only discussed a section of the design process of the rehabilitation robot, 138 presented a clinical appli- cation, and 16 included the design process plus a clinical application (Fig. 1 shows a flow diagram that illustrates the process of the selection of the papers).

The ten review articles examined a variety of rehabilitation robots for children

with physical disabilities. They were focused on children with neurological problems

(e.g. CP, ABI, and Stroke) or SCI and only investigated their use as part of physi-

cal therapy. In Fasoli et al. [35], Meyer et al. [61], and Bayon et al. [62], the robot

(6)

assistive therapies for children with CP were examined. Vova et  al. [63] reviewed the efficacy of functional electrical stimulation and exoskeletons in gait training to improve motor function and gait pattern in children with CP. Zwicker et al. [64]

reviewed the efficiency of robot-assisted treadmill training compared to traditional treadmill training in children with CP. Chen et al. [65] examined the effectiveness of various devices for upper limb robotic therapy on children with CP. The effects of robotic gait training practices in individuals with CP were investigated in Carvalho et al. [31]. Falzarano et al. [10] and Mahamud et al. [66] investigated upper and lower limb rehabilitation devices for neurological diseases. Dannenberg et  al. [67] com- pared different locomotor training, including robotic training, in children with SCI.

Compared with the previous reviews, this work analyses a broader range of aspects of paediatric rehabilitation robots, focused on the design parameters to fulfil the paediatric needs and how the technology and different conditions affect the robot design.

Paediatric robotic rehabilitation design requirements

Fifteen different requirements were identified (Table  2). The requirements are based on those proposed by Batavia and Hammer for assistive devices [68] and expanded by pro- posed requirements for paediatric rehabilitation devices highlighted by Weightman et al.

Fig. 1 Literature search flow diagram

(7)

[69], Bützer et al. [26] and Keller et al. [57]. In paediatric rehabilitation, it was apparent that the stakeholder’s needs related to operability, weight, safety, and motivation factors were relevant.

Table 2 Paediatric rehabilitation robots’ requirements and examples

Requirement Definition Example

Target group Range of ages and problem of the users ChARMin covered an age range from 5–18 years old [99]

Mechanical functionality The device performance, including the controlling level of assistance, the functional workspace, smoothness of movement and robustness

McDaid designed a gait trainer that allows children to stretch their legs through the entire ROM and support body weight up to 80kg [40]

Weight Total unsupported or unpowered mass of the device in relation to the user’s body weight

Lerner developed a Bowden cable structure for an ankle exoskeleton with a weight of 1.85 kg and placed 65% of the total mass above the waist to minimise the metabolic cost of walking due to the device’s weight [73]

Therapeutic benefit The type of exercise that the rehabilita- tion system should promote and how this will improve the user quality of life

The paediatric Anklebot provided inten- sive task-specific sensorimotor therapy to the ankle of children with motor disabili- ties to promote motor learning [75]

Safety The potential for the device to harm its

user IOTA device included a security stop

button that immediately halts the servo motors [175]

Comfort The user can use the device without

physical pain or discomfort The P-LEG robot used 3D printed braces based on 3D scans of the child’s legs to improve the child’s comfort [71]

Reliability The consistency of the device operation

in normal operating conditions Laubscher designed a gait guidance con- troller to guide the motion of the patient’s legs to follow healthy gait patterns to avoid unnatural gait patterns [176]

Operability The device is easy to control and adapt- able to changes in the user’s ability and sizes

ATLAS exoskeleton used a slide and tubular regulation size system to adapt to the fast growth of the patients at all stages [177]

Product appeal User satisfaction with the design, like fit,

appearance, and sound of the device One of the main requirements for PEXO was an appealing design, so the kidPexo version resembles a crocodile [26]

Quality of construction Typical use and care should cause no damage, distortion, or hinder the expected useful lifetime of the device

PEXO device did not have electronics in the hand module, making the device water and dustproof [26]

Social acceptability Matches user needs for discretion or

attention to avoid stigmatisation Weightman selected the handgrip of his robot through a questionnaire with different aspects like shape, style, feel, and colour [69]

Motivation Encompass any aspect of the device

considered to motivate the child ChARMin used an Audio-visual interface with various game-based virtual reality scenarios to motivate the child for active participation [57]

Cost The financial burden of the initial pur-

chase and ongoing costs of the device Volpini developed a low-cost robotic gait trainer to be used in developing countries [87]

Easy to maintain/repair The ease of keeping the device fully

operational, including when damaged P-Legs’ brace 3D print fabrication method made it easy to get new braces as the children grow [71]

Portability The possibility of the device to be trans-

ported between locations Cleary developed a smaller version of Pedbot that can be used at home [153]

(8)

In paediatric rehabilitation devices, operability is critical as children are in a continu- ous development phase during which their bodies, cognitive capabilities and physical abilities (e.g. skill levels) are changing, making them a “heterogeneous population” [52, 70]. Consequently, the device must adapt to different children’s abilities and sizes [71].

An important consideration is that the robot’s weight could obstruct the movement pattern of the limb and increase the child’s energy consumption [72, 73]. Furthermore, due to their musculoskeletal system’s immature development, their muscle strength and joint torque generation may not be adequate to assist in the movement being under- taken [74].

Concerning safety, children often cannot adequately assess the hazards of using com- plex technological devices [56]. Therefore, it is crucial to design safety mechanisms that minimise risky situations. These should be able to be activated remotely by adults with the child [57]. Furthermore, the use of compliant materials with shock-absorbing fea- tures (e.g. elastic elements like spring and Bowden cables) would be beneficial [17].

Finally, motivation is crucial because function recovery is not enough to engage chil- dren in the rehabilitation process [75]. Consequently, researchers have used strategies to engage children, like aesthetic designs attractive to the children [26, 76] or a virtual environment where they can interact with virtual objects [77, 78].

Type of robots used on paediatric robotic rehabilitation

Fifty-eight different devices were found that at least had a prototype in action. In Tables 3, 4, 5, 6, 7 (Figs. 2, 3, 4, 5, 6), the rehabilitation robots are presented chrono- logically and separated by their mechanical structure (end-effector or exoskeleton) and the anatomical part of the body where they are working (upper limb or lower limb). Furthermore, the tables summarise the characteristic features of the selected devices. This tabulated summary constitutes the reference for information provided in subsequent sections.

In respect of the developmental stage of the devices, the stages were classified into four categories: (1) commercial in the case the robots are available for its commerciali- sation; (2) clinical trial when the robot undergo a study where the participants were assigned to groups undergoing similar forms of therapy, but at different intensities, using various devices or undergoing various forms of therapy in a different order, aiming to determine the efficiency of therapy [28]; (3) feasibility study when the experiments con- ducted with a low number of people, often using the prototype of a device, to evaluate its safety and clinical feasibility without showing the potential benefits of the device [28];

(4) prototypes when the robots had not performed any test that involves people. It can be observed that 18 of them reached the commercialisation phase, but only 9 of them are certified by the US Food and Drug Administration (FDA). However, only 5 of the commercially available devices present a paediatric version of the rehabilitation system.

In the case of clinical studies, 34 systems conduct at least one clinical trial, 9 presented a feasibility study, and 15 are in the prototyping phase.

From the 58 devices, it was apparent that the majority (67%) were designed or had

been redesigned for children. When it comes to the type of robot, more than half were

an exoskeleton type structure. In the past five years, there is a trend (Tables 3, 4, 5, 6, 7)

(9)

for this structure to be more popular with designers than end-effectors for this structure robotic rehabilitation in paediatrics. Additionally, it can be noted that the majority of robots not explicitly designed for children are end-effector devices. In the case of the exoskeletons, the degrees of freedom (DOFs) are related to the number of joints and limbs that are powered. Therefore, it is possible to find exoskeletons with passive DOFs, which means that those joints are not actuated, but allow the free movement of the children´s joint. In contrast, for the end-effectors, the relation of DOFs of the robot and the actuated joints is not linear and depends on the robot’s mechanical design.

Actuators

Rehabilitation robots are moved by devices called actuators. Actuators convert a source of energy (e.g. electrical, thermal, pneumatic) into mechanical motion. Commonly reha- bilitation assisted robots are powered by electrical actuators. Among the compared sys- tems in Tables 3, 4, 5, 6, 7, over 93% of the robots used electric motors as the actuator, and about 7% used pneumatic actuators.

Table 3 Upper limb end-effectors rehabilitation robots

System

(year) Treated part of the body

DOFs Actuator Type of

rehabilitation Type of

training HCI input Paediatric disease (design for or treated condition)

Paediatric design Stage

of the device

Inmotion2/

Mitmanus (BIONIK, Canada) (1998) [126, 178]

Shoulder/

elbow 2 DC motors Physical

therapy Passive/

active/

assistive Imped-

ance Neurological No Clinical trial/com- mercial (FDA)

Wrist- Robot (2009) [119, 179]

Forearm/

wrist

3 DC motors Physical therapy

Passive/

active/

assistive Imped- ance

Neurological No Clinical trial

NJIT-RAVR (2009) [124, 180]

Shoulder/

elbow/

forearm

6 DC motors Physical

therapy Active/

assistive/

resistive Admit-

tance Neurological No Clinical trial GNO arm

(2009) [149]

Elbow 1 DC motor/

Cable driven

Assistance Assistive Finger move- ment

DMD Yes Feasibility

study AMADEO

(Tyromo- tion, Aus- tria) (2012) [150, 181]

Fingers 5 DC motors Physical

therapy Passive/

active/

assistive Imped-

ance Physical disabled children

No Clinical

trial/com- mercial (FDA) REAplan

(AXINESIS, France) (2012) [81, 182]

Shoulder/

elbow 2 DC motors Physical

therapy Passive/

active/

assistive

Position Neurological No Clinical trial/com- mercial

PASCAL (2013) [183, 184]

Shoulder/

elbow 3 Dc motors Physical

therapy Passive/

active/

assistive

Velocity Neurological Yes Clinical trial ReHaptic

(2014) [125, 185]

Forearm/

wrist

2 DC motors Physical therapy

Passive/

active/

assistive/

resistive Admit- tance

Neurological Yes Clinical trial

MyPam (2015) [166]

Shoulder/

elbow

2 Electric motors

Physical therapy

Active/

assistive

Position CP Yes Feasibility

study

(10)

Motor actuator

In robotic-assisted rehabilitation, the most common actuators are electrical motors with a rigid power transmission element such as a harmonic drive, ball-screws, timing belts, and chains. Unfortunately, their need for transmission negatively affects the back-driva- bility, efficiency, safety, size, and mass [97]. Nevertheless, they were likely chosen since they are efficient and easy to control. Some examples of paediatric robotic rehabilitation devices using electrical motors and rigid transmission are the Pedianklebot that used two brushless dc motors and a Rohlix linear traction device [98], the electric motor with

Table 4 Upper limb exoskeleton rehabilitation robots

System

(year) Treated part of the body

DOFs Actuator Type of

rehabilitation Type of training HCI

input Paediatric disease (design for or treated condition)

Paediatric design Stage

of the device

KINARM (KinArm, Canada) (1999) [82, 186]

Shoulder/

elbow

2 DC motors Physical

therapy

Passive/

active

– Neurological No Clinical

trial/

commer- cial IOTA (2013)

[175]

Thumb 2 DC

motors/

cable driven

Physical therapy

Passive/

active/

assistive Move- ment

Neurological Yes Proto- type

ChARMin

(2014) [57] Shoulder/

elbow/

wrist

6 Electric

motors Physical

therapy Passive/

active/

assistive Move-

ment Neurological Yes Feasibil- ity study Univer-

sidad Nacional de San Juan (2014) [130]

Elbow 1 DC Motor Physical

therapy

Passive/

assistive

EMG Injuries Yes Clinical

trial

Milwaukee University (2014) [187]

Wrist 2 actu- ated + 2 passives

DC motors/

cable driven

Physical therapy

Assistive Position CP Yes Proto-

type

GLOREHA (2016) (IDRO- GENET, Italy) [33, 107]

Hand 5 Pneumatic Physical

therapy

Passive/

active/

assistive Move- ment

Neurological No Clinical trial/

commer- cial (FDA) HAL

single joint (Cyber- dyne, Japan) (2019) [84]

Elbow 1 DC motor Physical

therapy/assis- tance

Assistive EMG CP No Clinical

trial/

commer- cial

PEXO

(2019) [26] Hand 2 Actu- ated + 1 passive

DC motors/

cable driven

Physical therapy/assis- tance

Passive/

assistive Push buttons or EMG

Neurological Yes Feasibil- ity study

PneuGlove (2019) [113]

Hand 5 Pneumatic Physical

therapy

Active/

assistive/

resistive Move-

ment CP Yes Feasibil-

ity study Exohand-2

(Android Technics, Russia) (2020) [131]

Hand 2 Each

hand

Electric motors

Physical therapy

Assistive EEG CP No Clinical

trial/

commer- cial

(11)

Table 5 Lower limb end-effectors rehabilitation robots

System

(year) Treated part of the body

DOFs Actuator Type of

rehabilitation Type of

training HCI input Paediatric disease (design for or treated condition)

Paediatric design Stage

of the device

GAIT trainer GT 1 (REHA-STIM MEDTEC, Switzerland) (2000) [188, 189]

Hip/

knee 2 DC Motors Physical

therapy Passive/

assistive Velocity Neurologi-

cal No Clinical

trial/

com- mercial

MOTOmed gracile (RECK, Germany) (2000) [190, 191]

Hip/

knee

2 DC motors Physical therapy

Passive / active/

assistive/

resistive

Velocity Physical disabled

Yes Clinical

trial/

com- mercial (FDA) IntelliStretch

(Rehabtek, USA) (2002) [192, 193]

Ankle 1 DC motor Physical

therapy

Passive/

active/

assistive/

resistive

Velocity and torque

Neurologi- cal

No Clinical

trial/

com- mercial (FDA) Innowalk

(Made for Movement, Norway) (2009) [114, 194]

Hip/

knee

2 Electric

motors

Physical therapy

Passive – Neuro-

muscular problems

Yes Clinical

trial/

com- mercial

National Taiwan Uni- versity (2009) [195]

Hip/

knee

2 DC motors Physical therapy

Passive – CP Yes Proto-

type

3DcaLT (2011) [134, 196]

Hip/

knee

4 Electric

motors/

cable driven

Physical therapy

Active / assistive/

resistive

Motion CP No Clinical

trial

Paediatric ICARE (2011) [197, 198]

Hip/

knee

2 DC motors Physical therapy

Active/

assistive/

resistive

Speed Neurologi- cal

Yes Feasi-

bility study Rutger ankle

CP system (2011) [112, 199]

Ankle 6 Pneumatic Physical therapy

Active/

resistive

– CP Yes Clinical

trial

SS-POINT (2013) [135, 200]

Ankle 2 DC motors Physical therapy

Passive/

active/

resistive

– Neurologi-

cal

No Clinical

trial TPAD (2014)

[102, 103] Hip/

knee 6 AC motors Physical

therapy Assistive/

resistive Motion CP No Clinical

trial Pedbot(2016)

[153, 201]

Ankle 3 DC motors Physical therapy

Active/

assistive/

resistive

Position CP Yes Clinical

trial Wyss Insti-

tute (2017) [202]

Hip/

knee

8 DC

motors/

cable driven

Physical therapy

Assistive Gait segmen- tation/

motion/

force

CP Yes Proto-

type

Pro-Gait (2017) [40]

Hip/

knee

2 DC motors Physical therapy

Passive CP Yes Proto-

type UFMG (2017)

[87] Hip/

knee 2 Electric

motors Physical

therapy Passive – CP Yes Proto-

type Leg Press

(2017) [89] Knee 2 Linear

electrical motors

Physical therapy

Assistive/

resistive

Imped- ance

Neurologi- cal

Yes Proto-

type

(12)

timing belt used in ChARMin [99], or the motor with chain transmission used in P.REX [100].

Some researchers [64] used a cable-driven transmission to replace the rigid transmis- sion for an elastic cable to improve the power to weight ratio and lower the inertia over the treated body segment. Examples include the ankle exoskeleton designed by The Uni- versity of Arizona [101] or TPAD [102], an end-effector robot for gait rehabilitation that used Bowden cables attached to the hip to generate assistive forces. This change in the transmission brings other advantages like modularity, simple architecture and is con- venient for reconfiguration, even though they present some disadvantages being unidi- rectional and difficult to model and control [97, 103, 104].

Following the concept of adding a flexible element in series with the actuator to improve the electric motors’ compliance, serial elastic actuators (SEAs) incorporate an elastic part in series with the electric actuator. This elastic element helps to decrease the actuator’s impedance and inertia and increases the back-drivability allowing better force control, even though they are limited by a large volume, heavy mass and complicated structure [17, 105]. An example of the use of this technology is the ATLAS exoskeleton [106].

Pneumatic actuators

Some authors considered that the mechanical linkage of the electric actuators is too heavy and can generate resistance at the joints, making them inadequate for rehabilita- tion applications [74, 107]. Instead of electrical motors, they used pneumatic actuators, consisting of a simple air pressurisation mechanism in an expandable chamber, convert- ing the energy from the compressed air to mechanical motion [18, 108, 109]. Their main advantage is improved back-drivability, and they are often lightweight at the site of actu- ation, have high power density, and can generate fast movements. They are not with- out limitations; firstly, poor portability because they need external compressors or fluid tanks as the power source. Secondly, it is challenging to create a good model and control strategy due to their nonlinear response to input pressure [109–111]. Among the devices analysed that used pneumatic actuators were the Rutgers ankle platform [112] for CP children and two gloves for hand rehabilitation PneuGlove [113] and Gloreha [33].

Training strategy

Devices for robotic rehabilitation may provide different training strategies depending on the type and severity of the patient’s impairment. These can be divided into passive, active, assistive, or resistive [42]. In general, the devices can offer more than one type of training.

In passive training, the force/motion is generated by the robot alone to perform the

exercise. The advantage of this training is that patients with minimal muscle activity

can receive therapy. For instance, through repetition of a movement, ROM can often

be maintained with muscles and joint structures (e.g. ligaments) repeatedly stretched,

ultimately maintaining their physiological length. Such movement reduces contractures

at joints, which can finally be very useful to caregivers making a notable difference to the

ease of transfers (e.g. sitting in a wheelchair to lying in bed). Examples of devices using

passive training are Innowalk [114] and Intellistretch [115].

(13)

Table 6 Lower limb exoskeleton rehabilitation robots

System

(year) Treated part of the body

DOFs Actuator Type of

rehabilitationType of training HCI

input Paediatric disease (design for or treated condition)

Paediatric design Stage

of the device

Lokomat (Hokoma, Switzer- land) (2007) [203, 204]

Hip/knee 4 DC Motors Physical

therapy Passive/

active/

assistive/

Imped-

ance Neurological Yes Clinical trial/

com- mercial (FDA) HAL

(Cyber- dyne, Japan) (2007) [91, 129]

Hip/

knee/

ankle

6 DC Motors Physical

therapy/

assistance

Assistive Foots- witch EMG

CP Yes Clinical

trial/

com- mercial (FDA) HWA

(Honda, Japan) (2007) [205, 206]

Hip 2 DC Motors Physical

therapy/

assistance

Assistive Move- ment

CP No Clinical

trial/

com- mercial University

of Verona (2011) [207]

Hip 2 Pneumatic Assistance Assistive Foots-

witch

CP Yes Feasibil-

ity study

Ekso-GT (ekso Bionics, USA) (2012) [127, 208]

Hip/knee 4 DC motors Physical therapy

Passive/

assistive Hip move- ment

ABI No Clinical

trial/

com- mercial (FDA) PediAnkle-

bot (2015) [75, 118]

Ankle 2

active + 1 passive

DC motors Physical therapy

Active/

assistive Imped- ance

Neurological Yes Clinical trial Walkbot

K (P&S Mechan- ics, South Korea) (2016) [209, 210]

Hip/

knee/

ankle

6 AC motors Physical

therapy Passive/

assistive/

active

Imped-

ance Physical

disabled Yes Clinical

trial/

com- mercial (FDA)

Robogait (Bama teknoloji, Turkey) (2017) [211, 212]

Hip/knee 4 Electric

motors

Physical therapy

Assistive Force Physical disabled

No Clinical trial/

com- mercial

WAKE-Up (2017) [17, 213]

Knee/

ankle

4 SEA Assistance Assistive Foots-

witch

CP Yes Feasibil-

ity study Univer-

sidad Pontificia Bolivariana (2017) [214]

Hip/knee 4 DC motors Physical therapy

Passive – Physical

disabled

Yes Proto-

type

CPWalker (2017) [92, 154]

Hip/knee 4 DC motors Physical therapy

Passive/

active/

assistive Imped- ance/

EEG/LRF

Neurological Yes Clinical trial Reha-

bilitation Institute of Chicago (2017) [94, 215]

Ankle 1 DC motor Physical

therapy Passive/

assistive/

resistive/

active

Torque/

position ABI No Clinical

trial

(14)

In the active training mode, the patient’s muscle can still generate activity on the affected limb. The robot does not help, making the patients perform the exercise by themselves at least partially. The active mode provides data concerning torques and the ROM produced, allowing assessments before and after therapy/surgery. For instance, Kinarm [116] and Lokomat [117] are devices that can perform active training.

For assistive or active-assistive training, the muscles of the affected body part can still be activated. Therefore, the patient can at least partially perform the exercise or

Table 6 (continued)

System

(year) Treated part of the body

DOFs Actuator Type of

rehabilitationType of training HCI

input Paediatric disease (design for or treated condition)

Paediatric design Stage

of the device

ATLAS (2017) [146, 167, 216]

Hip/

knee/

ankle

10 SEA Assistance Active/

assistive/

passive Foots- witch/

position/

force

SMA, SCI Yes Clinical

trial

P.REX (2017) [95, 100]

Knee 1 DC motor Physical

therapy/

assistance

Assistive Foots- witch/

position/

torque

CP Yes Clinical

trial

University of Arizona ankle (2018) [73]

Ankle 1 DC motors/

cable driven Physical therapy

Assistive/

resistive Foots- witch/

torque

CP Yes Clinical

trial

Tsukuba University (2018) [74]

Knee 2 Electric

brake

Assistance Assistive Foots- witch

CP Yes Feasibil-

ity study

Los Olivos University (2018) [217]

Hip/knee 4 DC motors Assistance Assistive Joystick DMD Yes Proto-

type

P-Legs (2019) [71]

Hip/

knee/

ankle

6 DC motors Physical

therapy/

assistance

Passive/

assistive Imped- ance

Neurological Yes Proto- type ExRoLEG

(2019) [218]

Knee 2 Linear

actuators

Physical therapy/

assistance

Assistive EMG/limit switch

CP Yes Proto-

type Cleveland

State University (2019) [176, 219]

Hip/knee 4 DC motors Physical therapy/

assistance

Assistive Position velocities

CP Yes Proto-

type

ExoRoboW- alker (2019) [220]

Hip/

knee/

ankle

6 DC motors Physical

therapy

Passive CP Yes Proto-

type Indian

Institute of Technol- ogy Guwahati (2020) [49]

Hip/

knee/

ankle

6 DC motors Physical

therapy

Passive Physical

disabled

Yes Proto-

type

Instituto Poli- técnico Nacional (2020) [221]

Hip/

knee/

ankle

6 DC motors Physical

therapy

Assistive EMG Physical disabled

Yes Proto-

type

(15)

movement without the robot. The assistance will be triggered after a particular event is detected through an HCI, allowing the patients to move further with the robot’s help.

Assistive training is relevant as it involves the active participation of the children. More- over, it improves the physiological responses needed to maintain and increase muscle strength and length, ultimately leading to improved ROM, in which the muscles provide some of the torque required. Due to these advantages, many designers have produced devices that use this training mode. Examples are Pedianklebot [118] and the wrist-robot [119].

As the name implies, the robot applies a force opposing the desired movement in resis- tive training, making the task more challenging. Resistive training is used to enhance muscle strength in the treated limb. This type of training was employed in the ankle device developed by the University of Arizona [120] and the upper limb end-effector NJIT-RAVR [121].

Human–computer interface (HCI)

The term HCI refers to methodologies to identify the user’s intent to move in the desired direction from different input sources and translate this intention into a command for the robot to move to facilitate the appropriate actions [122]. The designers who report upon the use of an HCI have primarily developed assistive training. Two main types of HCI inputs were identified: those associated with physical interactions and physiological signals [123]. In this aspect, the devices can rely upon only one signal as the input source or use two or more signals as input to start the desired movement.

The main physical interactions used on HCIs to control such robots are Imped- ance/admittance, body-powered control, and gait phases detectors. Impedance and

Table 7 End-effectors rehabilitation robots for upper and lower limbs

System

(year) Treated part of the body

DOFs Actuator Type of

rehabilitation Type of training HCI

input Paediatric disease (design for or treated)

Paediatric

design Stage of the device

KPT Cycla (Kinetec, France) (2010)(96)

Upper:

shoulder/

elbow Lower:

hip/knee

2 Electrical

motor Physical

therapy Passive/

active – DMD No Clinical

trial/com- mercial (discontin- ued)

Fig. 2 Picture of upper limb end-effectors rehabilitation robots: a Inmotion2/Mitmanus [79], b wrist robot [80], c REAPlan [81]

(16)

admittance control are the two most commonly used HCI. They are based on the rela- tion between position and force rather than controlling either force or position explic- itly. Impedance control accepts position or velocity as the input and outputs force or torque, and admittance is the opposite of impedance. Hence, force or torque are inputs, and velocity or position the outputs. This method could provide a natural, comforta- ble, and safe touch interface [122]. Some examples of devices that used this HCI are the NJIT-RAVR [124] and Rehaptic [125] upper limb robots that employed admittance con- trol or the robots for upper and lower limb Inmotion2 [126] and Pedianklebot [118] that applied impedance control.

When the children cannot generate an intention to move with the treated limb, body- powered control is applied. It consists of using the movement of a different body part as the trigger signal to initiate the rehabilitation robot. The main drawback of this approach is that it is hard to control many degrees of freedom due to the activation system’s sim- plicity. An example of this HCI is the Ekso robot, where the activation was made by moving one’s body weight laterally and then forward to trigger the assistance [127].

In assisted gait, a favoured approach for HCI is the use of gait phase detection. This technique identifies the different gait phases (heel strike, midstance, toe-off, and the swing phase) to apply forces to assist the children’s movement depending on the gait phase. Robots usually perform gait segmentation using inertial measurement units (IMUs) to detect angular velocities of the shank and/or the thigh, or footswitches

Fig. 3 Picture of upper limb exoskeletons rehabilitation robots a KINARM [82], b GLOREHA [83], c HAL single joint [84], and d PneuGlove [85]

(17)

to detect the foot’s ground reaction forces while the child is walking [128]. The main advantage is that splitting the gait cycle into discrete phases provides enhanced consist- ency and robustness to an inherently variable process and allows lower-level controllers’

implementation within each phase. The problem is that gait detection should be charac- terised for every target group, as the physical disability modifies the gait pattern [100].

An example of this type of system is the P.REX exoskeleton which utilised a combination of the footswitch and IMUs to detect the different gait phases to provide different levels of assistance within each phase [100].

Alternatively, for HCI based on physiological signals, Electromyograms (EMG) that measure electrical activity in the muscles and electroencephalograms (EEG), which measure electrical activity in the brain, are the main signals used. They are widely uti- lised because they can be obtained using non-invasive techniques without the need for medical intervention.

Concerning electromyograms, the primary type is surface electromyography (sEMG), a non-invasive and easy-to-configure procedure in which adhesive electrodes are placed on the skin above the muscle of interest. The benefit of using the EMG signal is that it allows detection of the user’s intent before the movement occurs. The electrical activity can be detected even if it is insufficient to generate movement of a joint. However, sEMG

Fig. 4 Picture of lower limb end-effectors rehabilitation robots: a Innowalk [86], b UFMG [87], c 3DCaLT [88], d Leg Press [89]. Reprinted from Biomedical Signal Processing and Control, Vol. 38, F. Chrif et al., Control design for a lower-limb paediatric therapy device using linear motor technology, Page 121, Copyright (2017), with permission from Elsevier

(18)

can suffer from contamination of the signal by electromagnetic interference, skin per- spiration, movement of electrodes and crosstalk artefacts. Also, for each muscle group of interest, a single EMG channel only shows the activation of that group. So, to perform an activity where many muscles fibres are recruited, it is necessary to use multichan- nel sEMG. Some examples of this technology in paediatric rehabilitation robots are the lower limb exoskeleton HAL [129] and the device for upper elbow rehabilitation of the San Juan National University [130].

The electroencephalogram (EEG) signal is recorded using many small surface elec- trodes, often configured in a bathing like cap placed over the scalp that detects the underlying electrical signals. The main advantage of the EEG signal is that the physical disability level does not limit it. Even if the patient has lost all their ability to move the

Fig. 5 Picture of lower limb exoskeletons rehabilitation robots: a Lokomat [90], b HAL [91], c CPWalker [92], d PediAnklebot [93], e wearable ankle rehabilitation robot developed by the Rehabilitation Institute of Chicago [94], f P.REX [95]

Fig. 6 Picture of KPT Cycla [96] an end-effectors rehabilitation robot for both a lower and b upper limbs

(19)

limb required for a task, the brain activity thought to be related to the intent to activate the muscles can be recorded. There are two main disadvantages to this system. Firstly, it is unsuitable for children with brain damage as they cannot generate standard brain patterns for limb activation. Secondly, the EEG signal has greater variability within it than the EMG signal, and it is also easily affected by changes in the patient’s mood and attention. Examples of the EEG signal use are the CP walker that used this signal as a part of its HCI to help children with a physical disability move their legs [120] and the Exohand-2 that used the EEG signal to interact with the exoskeleton [131].

Treated conditions

The majority of studies and devices were for children with neurological conditions (np = 183, 89%), CP being the most studied condition (np = 129, 63%). In contrast, other neurological disorders included ABI and strokes. Significantly few researchers investi- gated other conditions such as neuromuscular diseases (np = 15, 8%) and traumatic injuries to limbs and the spine (np = 6, 3%). The results obtained from the studies that perform clinical trials or a feasibility study suggests that robotic rehabilitation could benefit children with physical disabilities.

Neurological disorders

Concerning CP, there was evidence of improvement in physical disability using assisted rehabilitation robots. The benefits include an increase in muscle activity [120, 132], endurance for physical activities [133, 134], improvements of balance [114], walking speed [134, 135], the strength of the muscles [136, 137], ROM of the joints [84, 138], upper limb kinematics [139], and manual dexterity [33, 140].

For paediatric ABI, there were reports in the improvement of the walking ability [141], improvement of the lower limb motor performance [94], increase in the ROM of the wrist joint and force increase in the hand [142], improvement in motor function, and gait pattern [143, 144].

In children who suffer a stroke, three studies used rehabilitation robots while perform- ing physical therapy. Marini et al. [119] demonstrated an improvement in wrist motion after the robotic therapy, and Bützer et al. [26] showed the possibility of using a wear- able hand exoskeleton to assist children during task-oriented training could be helpful for rehabilitation therapies or assist children during ADLs.

Neuromuscular diseases

The neuromuscular disease presented a different scenario than neurological disorder due to the degeneration of their muscles as the disease progresses, making hard the use of rehabilitation robot due to stiffness in the robot’s joints, which can harm chil- dren’s weak muscles. Hence they require compliant actuation [145, 146]. Jansen et al.

[145] found that robotic rehabilitation therapy on upper and lower limbs help prevent

functional deterioration in children with DMD. Meanwhile, Ganguly et al. [147], Gar-

cia et al. [148] and Sanz-Merodio et al. [146] showed an improvement in walking ability

in children with SMA with the assistance of ATLAS, and the exoskeleton was designed

to provide Robotic-assisted gait training for children with SMA. Moreover, Koo et al.

(20)

[149] reported improved arm mobility in children with DMD while using a robotic arm device.

Traumatic injuries

Even if traumatic injuries are common in the paediatric population, robot rehabilitation has not been applied widely in injuries that differ from those at the head. Only scarce information was found related to these conditions. A study of hands robotic rehabilita- tion was found, highlighting the possibility of using robotic devices to treat burns [150].

Additionally, a study observed a significant improvement in the arm movement and elbowed angle after physical therapy using an upper-limb exoskeleton for 3 months in children that suffer a car accident [130]. Finally, another case reported improving walk- ing ability after robotic-assisted gait training in a girl with SCI [151].

Discussion

It is possible to see that various novel rehabilitation robots have become available to rehabilitation professionals and clients in recent years. And this trend will continue as is possible to incorporate them in activity programs aimed at improving independent func- tion [34–37] where they offer advantages over the traditional rehabilitation therapies, as they reduced the required effort of therapists during the exercises of the therapies, allowed massed practice in children with substantial limitations and provide informa- tion of the patient. Furthermore, they have the potential to be used as assistive devices to aid functional performance for users when they are worn. These possibilities will lead to a new variety of ways for assessment and intervention impacting users’ abilities, task demands, or the environment to promote functional performance and participation.

The findings of this review indicate that the design and development of robotic tech- nologies for the physical rehabilitation of children is in a preliminary stage of develop- ment, as many of the devices were designed for adult patients. However, there is a trend toward creating robots specifically for children [17, 26, 57, 148]. Yet less has been done to prove the benefits and constraints of such a system.

Traditionally, rehabilitation robot designers have focused solely on improving physical function [58], which can lead to rejection of the devices as not all the needs of children with disabilities are considered. Thus, to ensure successful adoption of the technology, the rehabilitation robots should cover these needs of the children. Hence, stakeholders’

cooperation is essential through their integration within the design and production pro- cess by providing feedback. Designers can use this feedback to validate that the robot meets the stakeholder’s needs. However, the fulfilment of these needs has strong rela- tionships to the chosen technology, mainly the type of robot, the actuator, the training strategy and the HCI. Thus, it is essential to know the advantages and disadvantages of the technology.

Type of robot

When it comes to the type of robot, we can see a trend to migrate from end-effector to

exoskeletons. However, most of the devices that had performed clinical trials were end-

effector robots designed for adults. This relation could be because the end-effector robot

(21)

works on the distal part of the limb, guiding the children limb through a movement [152]. This property is helpful in the case of operability as it does not require adaptation to match the children limb’s size, making it easy to be used by a diverse group of children.

Furthermore, the bulky frames over the patient limbs are avoided, helping to reduce the weight that the children need to handle. These advantages come with the problem of the systems requiring bulky and heavy external structures, reducing the device’s portability, constraining its use to medical facilities or specific spaces inside a building. Thus, limit- ing the amount of therapy that the children can have [153]. Additionally, the activity is restricted to a workspace constraining the number of possible movements [40], which could reduce motivation.

On the other hand, exoskeletons work in parallel to the patient limb to perform the activity. Hence, they can be portable devices with the possibility to provide assistive help during activities of ADLs and robotic-assisted rehabilitation therapy in a single device [26]. This advantage will help to provide free movement to enhance the subject’s motiva- tion and autonomously practise their movement training for longer periods [17, 154].

Furthermore, as technology advance, this freedom in mobility will help to increase the participation of children with physical disabilities in different social activities [155].

However, as the technology moves from clinical facilities to open spaces and robots interact more closely with the children, designers will face notable challenges (e.g. the irregularities of the surfaces on which one walks and how the robot reacts to perturba- tions outdoors environment). Consequently, the requirements of weight, comfort, safety, portability and social acceptability for the exoskeletons will be harder to achieve.

Actuators

The paediatric robotic rehabilitation technology is moving from end-effectors to exo- skeletons due to their versatility to be used as a rehabilitation tool or an assistive device [26]. Consequently, actuation technology starts to be a critical part of the design as it negatively influences the weight and the size of the robot.

For the end-effectors robots, actuators are not as critical as with the exoskeletons because they could be placed in external structures. This advantage makes it possible to use bulky and heavy actuators like electric motors. However, using electric motors is hard to achieve compliance that is an important property to increase safety as it is needed to avoid opposing forces that can injure the children. In end-effector robots, compliance was achieved using sensors and a control strategy [146] or using a soft mate- rial like the Bowden cables [103].

On the other hand, for exoskeletons, the robot design requirements are hard to enhance with the currently used technology. The actuation system components such as motor and rigid elements are designed for industrial applications not to interact with and to be worn by children. However, they are still the standard as they have the advan- tages of efficiency, are easy to control, and are readably available in the market. There- fore, the choice of the actuation system is crucial to improve the weight, portability and safety of the exoskeleton.

The first exoskeletons relied on electric motors with rigid transmissions, making them

bulky and heavy, reducing their compliance as they generate high resistive torque from

the metallic links of the exoskeleton. Therefore, making it difficult to move and less safe

(22)

can cause non-desirable inertial movements [156, 157]. Furthermore, they require exter- nal structures to manage the weight of the exoskeleton.

As the rehabilitation robots move from rehabilitation therapies inside a medical facil- ity to assist the children during ADLs, new actuation technology is needed. This chal- lenge led to using SEA actuators and cable transmission since they have the advantage of been intrinsically complaints as they incorporate soft materials, making the device safer.

Using Bowden cables in the transmissions brings other advantages like simple archi- tecture, low weight on the limb’s distal part, and easy to reconfigure. This last advantage is significant in paediatric rehabilitation as it allows to change the motor easily depend- ing on the abilities and size of the children [26, 73], even though they present some disadvantages because they become unidirectional and difficult to model and control.

Instead, in the case of SEAs, which still require rigid links, they were highlighted on the use for children with neuromuscular diseases, as some children are not only weak on the affected joint but the entire body. Thus the exoskeleton must hold the children, but at the same time being compliant to avoid inertial forces that can harm the weak muscles of the children [146].

Another type of actuator used on the robots was the pneumatic. Their attributes of low weight and easy-to-manufacture actuators of different shapes and sizes [158, 159] make them a desirable technology in this field. They are easy to adapt to children with various conditions. However, their main constraint is that they are typically connected to exter- nal mechanisms like compressors and pumps cumbersome and noisy. Thus, reducing their portability and appealing making them impractical to use outside clinical facilities.

In Table  8, the advantages and disadvantages of the current actuator technology are presented. It is possible to notice that there is no perfect actuator technology, so more research in this area is needed. Moreover, in the future will be interesting to see devices that use different soft actuators technologies that are inherent compliant and light- weight, such as the already mention SEAs, pneumatic, and Bowden cables. But also new technologies that are under research to be used on rehabilitation robots, like shape

Table 8 General summary of advantages and drawbacks of each actuation technology

Actuation technology Advantages Drawbacks

Electric motors High precision Easy to control

Readably available in the market

Not compliant Large size Heavy Noisy

SEA High precision

Easy to control Compliant Better force control

Large size Heavy

Complex structure

Bowden cable Modularity

Simple architecture Easy to reconfigure

Low weight on the distal part of the limb

Unidirectional

Difficult to model and control

Pneumatic Lightweight

Compliant

Have high power density Fast actuation

Low cost

Easy to manufacture in different shapes and sizes

Poor portability because they require external components Difficult to model and control

(23)

memory alloys [160], dielectric elastomer [161], or twisted and coiled polymers actua- tors [162], as they will reduce the overall weight and increase the compliance. Further- more, this new technology can be manufactured in different sizes and shapes [163] that could be easily adapted to robots for children of varying height and ability conditions.

Training strategy

In the case of the training strategy, there is no best strategy, but it rather depends on many factors like the abilities and disease that the children have. For example, passive training is suitable for patients with limited mobility; however, when the children are able to generate movements, it tends to decrease children’s participation during the exercise, thus reducing the efficiency of the training [164]. That is why most of the research on the training strategy is centred on assistive training, where the children’s participation is needed. This engagement with the therapy increases the motivation of the children to perform the activities, enhancing the benefits from the therapy [165].

Another advantage of assistive training is that it is used together with video games to increase children’s motivation and social interaction [166]. In addition, this strategy is required for assistive devices. It needs to provide the required intensity to generate the movement safely, efficiently, and reliably, depending on the applied force by the user [167]. However, there is no clear which is the best strategy to provide assistive movement, where some examples of different assistive strategies are guidance force, path control, and locomotion strategy [57, 154, 165, 166].

Alternatively, some researchers suggest that resistive training could be more benefi- cial for rehabilitation therapy than an assistive force, as it increases the engagement of the children, which can help drive motor learning [120, 168]. Hence, further research is required on the optimal training strategy to increase the benefits from the rehabili- tation therapies.

Human–computer interface

The HCIs are essential in developing robotic rehabilitation robots, as they are the medium for the interaction between the children and the robot, impacting the function- ing of the actuators and training strategies directly. Thus, HICs are a crucial factor for safety and motivation needs, as it is how the children “communicate” their intention to the robot. Consequently, If the HCI is complex to use, it could lead to the rejection of the device [52]. Furthermore, for safety reasons, the HCI must detect the trigger signal properly and discern between intended movements and involuntary movements, as it can generate undesirable responses [41, 169]. For instance, in the case of CP patients presenting increased muscle tone, rapidly occurring muscle spasms, and severe jerks, which can be considered as a deviation of pattern, causing the robot to apply undesired forces to correct for it or turn off the device suddenly [170].

Consequently, selecting the best HCI for every case needs to be evaluated depending

on the capabilities of the children. For example, it could be challenging for patients with

advanced muscular dystrophy to use EMG and admittance/impedance interfaces as their

muscles progressively deteriorate, turning unable to activate the muscles to generate a

(24)

movement or a detectable signal [149, 171]. On the other hand, EEG could lead to a better motor function recovery for children with CP as it integrates the central nerv- ous system into rehabilitation therapy [154]. However, this technology could be hard to implement in patients with a cognitive deficit, requiring concentration [131, 172]. For HCIs, it would be interesting to see more devices using physiological signals as they can also evaluate the efficiency of the therapy [154], novel approaches of body-powered con- trol to address children with limited mobility of limbs [173], and devices that integrate different HCI strategies to make the system more robust and adaptable [71, 167, 174].

Treated condition

To better understand how the technology can positively impact children’s lives, it is essential to analyse how the rehabilitation robot’s technology has addressed the different paediatric conditions that can generate physical disability. Because, even if they share in common the deterioration of the musculoskeletal system, each one of the conditions presents certain specific characteristics that need to be considered.

Most of the research has focused on children with neurological conditions, particu- larly children with CP. However, it seems likely that many of the designed robots that currently work with neurological conditions could also be utilised in traumatic injury scenarios, especially because they have been designed for rehabilitation therapies that can improve common problems across both neurological and traumatic injuries like ROM deficits and a lack of ability to generate muscle force [80, 150]. Contrarily, neuro- muscular diseases present a different scenario than neurological disorders and traumatic injuries as the diseases are progressive, making it hard to obtain a permanent improve- ment on the children skills. Thus, the designs have been focused on design devices capa- ble of assisting with exercise and helping with the ADLs to maintain specific abilities (e.g. walking) for a longer period of time [122, 155].

Unfortunately, the outcomes proving the efficiency of rehabilitation robot in chil- dren is still scarce, as the pieces of evidence are low and weak. Hence, the information coming from these studies should be asses carefully, as there are very few randomised controlled trials, with small sample sizes and variability in children’s ability, outcomes measures, treatment protocols, and used devices [31, 65]. Thus, to better understand if the designed robots fulfil the paediatric needs properly by improving their quality of life and physical ability, more studies and robots designed especially for them are needed.

Furthermore, more studies with children presenting different conditions from neurolog- ical ones are needed, as it can be noticed that the treated condition impacts the require- ments design of the rehabilitation robot.

Limitations

It was apparent that some studies were conducted with participants from a wide range

of ages; therefore, it was difficult to target all the articles that include paediatric partici-

pants. Another problem was the upper bound on the paediatric population’s age as some

papers with the term young adults included paediatric participants.

Referenzen

ÄHNLICHE DOKUMENTE

 human issues: method application within an interdisciplinary team of engineers and user of the medical device; involvement of the primary user in particular

Moshaii, A.A., Moghaddam, M.M., Niestanak, V.D.: Analytical model of hand phalanges desired trajectory for rehabilitation and design a sliding mode controller based on this

With the prototypes and tools developed in this thesis I have shown that it is feasible to design and control a human-inspired robotic eye that fea- tures human-like performance and

1,9,17 A Turkish study performed by Almaz et al 1 was the first randomised controlled clinical trial reported on the reduction of caries by licorice lollipops with the

The exclusion criteria were: studies that assessed caries experience in children with systemic diseases; outcomes based on ques- tionnaires applied to the caregivers; association

kinematic analysis, mechanical design, control development and experimental evaluation of two test rigs for generation of various stepping movements in a supine-lying position

We optimized the number of used PCs and the feature weighting factors using non- linear optimization techniques and exhaustive search [see Section V-C] on the basis of a large set

In order to provide native data-level interoperability between different robotic frame- works we have developed a generic approach to mediate between the different technolo- gies,