• Keine Ergebnisse gefunden

2.3.1 Density

To calculate molar concentrations,c, the densities, ρ, of the investigated systems were de-termined with a vibrating tube density meter (Anton Paar DMA 60, DMA 601HT) accord-ing to the method of Kratkyet al.118The temperature was kept constant to±0.003C with an overall accuracy of±0.02C using a Braun Thermomix 1480 thermostat in combination with a thermostated heat sink (Lauda RK 20). The instrument was calibrated with de-gassed water (Millipore MILLI-Q) and purified nitrogen at atmospheric pressure, assuming densities from standard sources.119 The precision of the measurements was ±0.001g L1. Taking into account all sources of error (calibration, measurement, purity of materials), the overall uncertainty ofρ is estimated to be within ±0.05g L1.

In addition, the densities of selected [emim][EtSO4] + AN mixtures were measured with a density meter DMA 5000 M (Anton Paar GmbH, Graz, Austria). The temperature was kept constant to ±0.001C with an overall accuracy of ±0.01C with a built-in

thermo-†We thank C. Schöggl-Wagner and T. Feischl, Anton Paar GmbH, Graz, Austria, for performing the measurements.

electric temperature control. The precision of the measurements was ±0.001g L1. The accuracy stated by the manufacturer is ±0.005g L1, although real errors are certainly higher.

The data obtained for selected [emim][EtSO4] + AN mixtures using the DMA 5000 M density meter are given in Appendix A.1.

2.3.2 Conductivity

Electrical conductivities, κ, were determined with the equipment described by Barthel and co-workers,120,121 consisting of a home-built precision thermostat stable to <0.003K in combination with a thermostated heat sink (Lauda Kryomat K 90 SW). Additionally, a new homemade precision thermostat was set up in combination with a thermostated heat sink (Julabo FP40, flow rate tunable via a needle valve) for measurements at high temperatures. Within slight variations, the setup corresponds to that shown in Figure 1 of ref. 122122. High-temperature stable materials were used throughout. This thermostat was stable to < 0.003C over the entire temperature range. With the currently used silicon oil (M50), it is limited to 215C. The resistance of the temperature bridge, Rt, was calibrated with the use of a NIST-traceable Pt sensor and bridge (ASL). The values of Rt together with all relevant parameters required for the temperature setting are listed in Table 2.2. An adequate interpolation formula is given by

Rt/Ω = 470.53

1 + 3.9836·103t/C5.9931·107(t/C)2

(2.26) One set of three-electrode and two sets of two-electrode capillary cells were used depending on the values of κ. The cells with cell constants C in the range of (25 to 470) m1, (12 to 46) cm1 and (25 to 360) cm1, respectively, were calibrated with aqueous KCl.123

The cell resistance, R(ν), was measured with a manually balanced high-precision conduc-tivity bridge as a function of AC frequency, ν, between 480Hz and 10kHz. To eliminate electrode polarization, the measured resistances were extrapolated to infinite frequency, R = limν→∞R(ν), using the empirical function R(ν) = R+A/νa; the parameter A was cell-specific, and the exponent a was found to be in the range 0.5 a 1.124 The conductivity was obtained asκ=C/R.

The uncertainty in temperature was ±0.01C. Repeat measurements of selected samples with different cells agreed within ±0.5 %. Thus, this value may be taken as an estimate for the relative uncertainty of the obtained values.22,125 The effect of temperature on the cell constants C was well below the stated accuracy.125

The setup described was used to determine precise conductivities of binary IL + AN mix-tures at 25C22and of neat imidazolium-based ILs as a function of temperature.125Reprints of these recently published papers are given in Appendices A.2 and A.3.

2.3.3 Viscosity

Viscosities,η, of [emim][BF4]#2 + [emim][DCA] mixtures (Section 4) were measured under an argon atmosphere at (25±1)C using a CVO 120 high-resolution rotational viscometer

Table 2.2: Parameters for the temperature settings of the high-temperature thermostat:

temperature of the cold source,θcs, opening of the needle valve,onv (0 =closed), resistance of the temperature bridge,Rt, temperature displayed at the ASL, θASL, and PID controller settings (proportional, P, integral, I, and derivative, D, values together with integral, τI, and derivative,τD, time constants).

θcs/C onv / rotations Rt θASL/C P I τI D τD

10 5.0 517.22 24.996 3 8 6 5 4

25 4.5 535.79 35.006 4 6 6 5 4

35 4.0 554.31 45.004 4 6 5 5 4

45 3.5 572.77 55.005 4 6 6 5 4

55 3.0 591.18 65.001 4 6 6 5 4

65 2.5 609.51 75.002 4 6 6 5 4

75 2.0 627.81 85.004 4 6 5 7 4

85 1.5 646.05 95.004 3 6 5 7 4

95 1.5 664.23 105.000 3 6 6 7 4

105 1.0 682.37 115.002 4 6 7 7 4

115 0.5 700.43 125.001 5 7 7 7 4

- 0 718.44 135.000 5 7 7 7 4

- 0 736.38 144.998 5 7 7 7 4

- 0 754.28 154.996 6 7 7 7 4

- 0 772.13 164.999 3 7 7 9 4

- 0 789.92 175.005 3 7 8 9 4

- 0 807.63 184.998 3 7 7 9 4

- 0 825.31 195.002 4 7 7 9 4

(Bohlin Instruments, UK) and a cone of 4 slope with 40mm diameter. The instrument was calibrated by calculation of a correction factor,fcorr=ηLit/η, using averaged published data, ηLit, for [emim][BF4].126–128 A similar value for fcorr, albeit with a much larger un-certainty, was obtained for neat [emim][DCA] but was not considered further because the published viscosities scattered considerably. The corrected viscosities were than obtained as ηcorr =fcorr·η. Repeat measurements indicate a precision of ±2 %. This value may be taken as an estimate for the uncertainty of the measurement.

Additional measurements of selected [emim][EtSO4] + AN mixtures were performed using a AMVn automated micro falling ball viscometer (Anton Paar GmbH, Graz, Austria) with built-in thermoelectric temperature control (0.01C resolution; <0.05C accuracy). The instrument was equipped with capillaries of varying diameter, dcap = 1.6 or 1.8mm, and fitting balls calibrated by the supplier. The diameter of the capillary was chosen with respect to an optimized rolling time of the ball. With these two capillaries measurements in the range 0.3≤η/mPa s≤70are realizable. The precision stated by the manufacturer is better than 0.5 %.

The data obtained for selected [emim][EtSO4] + AN mixtures using the AMVn viscometer are listed in Appendix A.1.

2.3.4 Refractive indices

Refractive indices, n, of [emim][EtSO4] + AN mixtures were measured at 25C using an Abbemat WR MW automatic digital refractometer (Anton Paar GmbH, Graz, Austria), equipped with a custom-made, variable light source made from LEDs (tunable to 437.0, 488.1, 515.0, 531.8, 589.2 and 632.2 nm). The instrument measured the critical angle of total reflection by shadowline detection with a CCD array. The temperature was adjusted with a built-in Peltier thermostat, with a stated stability of ±0.002C and accuracy of

±0.03C. As specified by the manufacturer, the resolution and overall accuracy of the measurement were 1· 106 and 4· 105, respectively. Repeat measurements were well within the latter range.

The data obtained for [emim][EtSO4] + AN mixtures are listed in Appendix A.1.