19 SCIENTIFIC INSTRUMENTATION 19.1 Surfmet

The Surfmet system, which combines the old Met and TSG systems ran continuously for the duration of the cruise with data logged to level B and also sent to the OTD Met system via a serial link.

The remote temperature sensor measuring incoming non-toxic water temperature was suspected of jumping and drifting. This was replaced with a spare but this too was found to jump at certain times. This may be attributable to the physical properties of the non-toxic system which may cause some heat generation/loss whilst on/off station. It was not always apparent though and requires further observation.

Prior to cruise D232 a new non-toxic pipe system was installed. This is plastic coated piping and there is a direct feed to the TSG flow-through system. The old header tank is now replaced by a vortex debubbler which operates at 40-50 1 min-1. with small volume, thus reducing lag time. A flow-through transmissometer and fluorometer are fed from the same supply as the TSG.

The output from the TSG was modified to provide an output to the CO2

measuring equipment, although flow to the TSG was reduced it appears to have had no detrimental effect.

The windvane of the Met system is oriented so that zero degrees is to Port. For this cruise however, the crossarm which supports it and the anemometer was rotated so that zero degrees was forward.

19.2 ADCP

The previous cruise showed that although one of the transducers four beams was defective, the ADCP could still operate using three beams. At first the data appeared to be good but halfway into the cruise, the defective third beam appeared to be producing some bad signals. This meant that data signals of bins deeper than 200 m were corrupted. The third beam signals were then grounded at the receiver board in the deck unit and the problem was resolved. Data down to 400 m then appeared to be good and matched closely with the LADCP which was being used on the cruise.

19.3 EchoSounder

During the early part of the cruise the echosounder suffered from considerable noise. This meant that there were a lot of drop outs and false depths given on the digital output, although it was still possible to see what the depth was from the scrolling display. About two weeks into the cruise this noise seemed to disappear but was replaced by weak signals, which also produces a lot of depth errors. This problem was less apparent in depths less than 2500 m where the soundings were consistently good. The problem appeared to be with the transmission from the deck unit but since the latter part of the cruise was shallow no further investigations were carried out.

19.4 SBWR

The system was reinstalled prior to this cruise after calibration and fitting of valves to the inlets. A fault was found with the Port Pressure transducer but this was eventually traced to a broken wire in the signal circuit. The SBWR ran continuously throughout the cruise with a change of sampling parameters midway through in order to optimise the statistical analysis.

19.5 XBT

The XBT system was used to deploy about 35 probes, including T5s and T7s in the latter part of the cruise. The launching and data collection worked fine but the GOES transmitter buffer was full and didn't empty at the scheduled transmission time.

Dave Jolly 20 SCIENTIFIC ENGINEERING

Cruise 233 consisted of a 139 CTD deployments through the starboard gantry, using the 20 ton Cobra winch system and the 10 ton Cobra winch system. Also in use were the Non toxic systems and Milii pore water plant. A few minor problems occurred during the cruise but none led to any major loss of equipment or scientific down time.

20.1 Starboard Gantry

The gantry worked well and caused no problems throughout the cruise.

20.2 20 ton Cobra winch system

This system was used with the deep tow electrical conducting wire for the deepest casts. There were no problems with this system after the initial setting up of the back tension loads on the storage drum. Trials were undertaken by RVS technicians to try to determine an intermittent fault with one of the boost pumps, however this did not interfere with the scientific cruise programme.

20.3 10 ton Cobra winch system

This system was used for the majority of the casts and in general worked well.

There were, however, a few small problems.

Winch spooling

The recovery of the wire had to be slowed on a few occasions to help to prevent wire distortion. This problem seemed to cure itself after a few deep casts and there were no more problems encountered.

Diverter sheave bearing

One of the inboard sheaves bearings collapsed and needed repair. These repairs were undertaken by the RVS technicians. A new bearing was turned on the lathe, fitted, and the unit reassembled. The sheave gave no more problems.

Retermination of the CTD wire

An electrical fault on the termination was found. The wire was cropped at 135 m and reterminated. After being load tested the wire was put into service and gave no more problems.

20.4 Non toxic

A few leaks followed the refit modifications but no serious problems occurred.

20.5 Milii Q water plant

The system was serviced by RVS technicians and a circuit board replaced, no major problems were encountered.