• Keine Ergebnisse gefunden

5. GMR-TYPE MAGNETIC BIOSENSOR

5.3. OOMMF model

Chapter 5: GMR-type magnetic biosensor

-15 -10 -5 0 5 10 15

0 1 2 3 4 5 6 7

90°

GMR-amplitude [%]

magnetic field [kA/m]

Figure 42: Isotropic magnetoresistance response of a spiral-shaped sensor element

we choose the easy direction (field parallel to the long side) and the hard direction (field parallel to the short side) of our model system. Furthermore, we implement an additional simplification by simulating only two magnetic layers separated by a single non-magnetic spacer layer. Such a tri-layer model system produces the same GMR response like a multilayer stack if the effective af-coupling is doubled for simulation purposes (in an infinite multilayer system, each magnetic layer experiences af-coupling from two adjacent ferromagnetic layers).

A number of parameters enter the micromagnetic simulation. Straightforward are the thickness of the ferromagnetic layers (1.6 nm) and the saturation magnetization of Py (860 kA/m at room temperature, Ref. 143). The cellsize is set to 20 x 20 nm2, which corresponds to the size of a typical grain in a Py-layer. The standard six-neighbor exchange energy term is taken for the ferromagnetic coupling, and the exchange coefficient is set to its full value for Py (13 pJ/m, Ref. 143). Due to the reduction of our actual multilayer system to a tri-layer, the antiferromagnetic coupling strength is set to 3.8 µJ/m2, which is twice the value derived in chapter 5.2. Further energy terms are the demagnetizing energy and the Zeeman energy, which can include both a spatially homogeneous external field of adjustable strength and direction and various dipole fields from magnetic particles. Even though Py does show a small crystalline anisotropy of KV = -0.4 kA/m (Ref. 143), it is not included in this simulation due to its negligible magnitude and the fact that our magnetic layers are polycrystalline. In terms of simulation parameters, a total torque across all spins of the system of dm/dt

= 0.01 °/ns is chosen as a stopping criteria at each stage.

-16 -12 -8 -4 0 4 8 12 16

-600 -500 -400 -300 -200 -100 0 100 200 300 400 500 600

MX [kA/m]

HX [kA/m]

length = 1 µm length = 2 µm length = 4 µm length = 8 µm

Figure 44: Projection of the total magnetization onto the long axis as a function of applied external in-plane field for different lengths of the model system

Another set of parameters comes from the dimensions of the line which is used as a model system in this simulation. According to the respective value of the spiral windings, its width is set to 1µm. Its length, however, is not determined that clearly.

Looking at the mid-section of a spiral-shaped sensor element shown in Figure 43, one could say that a length of 4 µm approximates the tangential of a winding without too much error due to the neglected curvature. Another criteria is the shape anisotropy of the model system, which should be large in order to correctly resemble the situation within a spiral winding. Figure 44 shows the simulated field dependence of the total magnetization for different lengths of the model system. The field is applied in the easy axis (x-direction), and the respective magnetization component is calculated relative to the total volume of the model system (the spacer layer is set to

Chapter 5: GMR-type magnetic biosensor

the same thickness as the magnetic layers, which is why the displayed saturation magnetization is only 2/3rd of the Py-value). To increase computation speed, the cellsize is set to 200 nm for this test only. As expected, the saturation field becomes smaller with increasing shape anisotropy, but the difference between a length of 4 µm and 8 µm is only marginal. Thus, a stripe length of 4 µm is chosen to save some computation time.

No periodic boundary conditions are applied at either edge of the model system, which results in stray fields and diverted magnetization patterns at its borders. In a real spiral segment, this is not the case along the easy axis, since the windings are similar to closed loops. However, the oommf code is not designed for implementing periodic boundary conditions, so that additional edge effects have to be kept in mind when comparing simulated results to measurements.

Layer 1

MG1mn

MG2mn

Layer 2

x

(direction of current) y

Figure 45: Sketch of the relative resistance calculation for the model system

The main output of the oommf simulation program is the local magnetization vector at every cell for each value of a stepped external homogeneous magnetic field. In order to compare these simulated results to measurements, the magnetization information has to be converted into magnetoresistance curves, i.e. relative electrical resistance over applied magnetic field. Since the current flows parallel to the spiral windings, it is taken to be oriented along the easy axis (x-axis, see Figure 45) of our model system.

In order to calculate the total resistance of the model system, the relative resistances of all pairs of opposing cells are computed from the angles between their respective magnetization vectors. For cell number m,n in this two-dimensional array, its resistance rmn is given by:

( )

1mn 2mn

mn 0 mn mn 2

M M

r r 1 A 1 cos with

2 M

  ⋅

=  + − α  α =

G G

G Equation 17

Here, αmn is the angle between the two magnetization vectors MG1mn

and MG 2mn, r

0 is the cell resistance in saturation (at αmn=0°), and A is the full GMR amplitude. Next, the relative resistances of all possible conduction paths are calculated by summing the individual cell resistances along the x-axis (total number of M cells in each path).

These conduction paths are parallel to each other, so that the total resistance R of 63

the model system for a particular magnetization configuration is obtained by taking the inverse sum of all conduction paths along the y-axis (N conduction paths).

Relative to its saturation resistance R0, it has the following form:

(

mn 1

n m

0

R N A

M 1 cos

R =M + 2 α

)

 

∑ ∑

Equation 18

These calculations are carried out for each step of the external field, resulting in complete GMR curves that can be compared to the data. Strictly speaking, the construction of the relative resistance from the magnetization configuration is not entirely accurate, since in this model, there is no possibility of a current flowing in a diagonal direction. However, such paths have a larger total resistance and only contribute in a negligible way.

-35 -30 -25 -20 -15 -10 -5 0 5 10 15 20 25 30 35 1,00

1,01 1,02 1,03 1,04 1,05 1,06

1,07 measurement averagedsimulation:

full af-coupling half af-coupling

Rrel

in-plane magnetic field [kA/m]

-35 -30 -25 -20 -15 -10 -5 0 5 10 15 20 25 30 35 1,00

1,01 1,02 1,03 1,04 1,05 1,06

1,07 easy axis (x) hard axis (y)

Rrel

in-plane magnetic field [kA/m]

(a) (b)

Figure 46: Relative magnetoresistance of the model system

a) simulated GMR characteristics along the hard and easy axis of the model system b) comparison of average simulated and measured characteristics

Figure 46 displays GMR characteristics obtained from simulations of the described model system. The amplitude A is set to 7.4 %, which corresponds to the data taken for a real spiral-shaped sensor element. In part a), the individual magnetoresistance curves along the x- and y-axis are shown. Clearly, the saturation field is smaller when the field is applied in the easy direction. The reason why the amplitude does not reach its full value of 1.074 lies in the stray fields at the borders of the model system.

These edge effects prevent a perfectly antiparallel alignment of the magnetization vectors of the two magnetic layers.

In Figure 46 b), the two individual simulated GMR characteristics are averaged and compared to a measured magnetoresistance curve of a spiral-shaped sensor element. Clearly, there is quite some disagreement in the saturation field, the behavior at small fields (the resistance is almost stationary in the simulated case) and the hysteresis. Such differences are to be expected due to the oversimplified model and the perfectly straight edges of the simulated rectangle, which tends to produce symmetric vortices around zero external field. These vortices are the reason for the almost constant characteristic at small fields, since they are pretty stable and need a comparably large field to break up.

Most of the parameters that enter the simulation are pretty fixed and cannot be varied without leaving the physical basis. The only exception is the ferromagnetic coupling, which could be smaller at the edges of grains due to crystal mismatches. However,

Chapter 5: GMR-type magnetic biosensor

reducing the exchange coefficient does not affect the simulated GMR curves dramatically, and no better agreement to the measured characteristic can be obtained this way. Thus, it is left at its original value of 13 pJ/m. The saturation field mostly depends on the antiferromagnetic coupling coefficient, and a reduction to one half of its original value (1.9 instead of 3.8 µJ/m2) results in a much better agreement in this respect. Even though such a modification cannot be justified within the physical picture, it is undertaken for all subsequent GMR type oommf simulations in order to improve their quantitative quality. The values of all other parameters remain like they are stated above for all following oommf calculations.

Despite the adjustment of the af-coupling, the characteristics are still quite different at small fields, which effectively disables real quantitative comparisons to the measurements described below. However, the simulations within this model system are still valuable, since they help in interpreting the measurements qualitatively.