• Keine Ergebnisse gefunden

Calculation of the pπ 0 η cross section

Data and data analysis

3.6 Calculation of the pπ 0 η cross section

0 0.2 0.4 0.6 0.8 1 0

5000 10000 15000 20000 25000 30000

γ Σ η π0 x +

p γp CL of

Figure 3.6: Confidence level for the Monte Carlo simulations of the reaction γp→pπ0η

projected on the inner detector9 using θpf it and the Z-components are deduced: Zmin and Zmax. The Z component of the proton from the fit should be between these two values. Moreover the Z-component of all photons should lay outside this projection or

∆φ of the photon and the proton from the fit should be larger than 20.

As the result 18,791γp→ 0η and 137,204 γp→pπ0π0 events were reconstructed. Under theη peak there is a background estimated to (9±2)% of the signal events. The background was obtained from side bins of the η peak in fig. 3.8. For each event from a side bin, the data event closest in phase space was found and subtracted as a background event. The corresponding invariant mass and angular distributions of events from side bins and the correspondent events from the data sample were checked and found to coincide within error bars.

0 200 400 600 800 1000 1200 1400 200 0

600400 1000800

140001200 200 400 600 800 1000 1200 1400

x102

π π

0 0

π η

0

π η

0

m

m γγ

γγ

Figure 3.7: The invariant gamma-gamma mass of the first two photons is plotted against the invariant gamma-gamma mass of two other photons. A clear signal from 0π0 events and two bumps resulting from 0η events can be seen.

plane10 in the center of mass system, invariant mass of 0 (m0) and invariant mass of (m) have been chosen. These variables are labeled with indices i,j,k,l,mrespectively.

The photon is calculated from the virtual wire chamber. The total cross section for a given energy intervala is given by

σtota = 1

ρtarget ·BRπ0γγ·BRηγγ·εtrigger ·

i1

i=i1−N

1

Nif lux ·

j,k,l,m

Ni,j,k,l,mrec data

εM Ci,j,k,l,m (3.18) εM Ci,j,k,l,m= Ni,j,k,l,mrec M C

Ni,j,k,l,mgen M C,

whereNif lux is the flux per wire,Ni,j,k,l,mrec data is the number of reconstructed events after all cuts in a corresponding five dimensional bin,Ni,j,k,l,mrec M C is the number of reconstructed Monte Carlo events in the same five dimensional bin, and Ni,j,k,l,mgen M C represents the number of generated

10The momentum vectors of the three particles in the final state define the decay plane in the center of mass system.

[MeV/c]

γ

Mγ

100 200 300 400 500 600 700 800 900 1000

Events

1 10 102 103 104

>10%

γ γ π0

CLp

<1%

η π0

>10% and CLp γ

γ π0

CLp

<1%

π0

π0

>10% and CLp γ

γ π0

CLp

=150048

π0

π0

N

p

=20416

η π0

N

p

Figure 3.8: Illustration of the confidence level selection for the γp →pπ0π0 and γp→pπ0η reactions. The black curve shows the events after requiring CL0γγ > 0.1.

The red curve shows which background will be rejected if both cuts are made:

CL0γγ >0.1 andCL0η >0.01. The blue curve shows the rejected background for CL0π0 >0.01 and CL0γγ >0.1.

Monte Carlo events in the aforementioned five dimensional bin. The ratio of Ni,j,k,l,mrec M C to Ni,j,k,l,mgen M C is called acceptance εM Ci,j,k,l,m. ρtarget = 2.231 · 10−7µb−1 is a surface density of the liquid hydrogen target and εtrigger is a trigger efficiency. The number of reconstructed events, corrected for their detection and reconstruction efficiency εM C, is summed over all wires pertaining to the energy bin. The photon flux for this energy bin is also obtained by summation.

The total cross section forγp→pπ0ηis reconstructed from measured data (see fig. 3.9) with a 20 MeV bin width in

s, due to resolution of the upper proportional chamber of 17-20 MeV.

Events due toγp→pπ0π0 are also reconstructed by using almost the same selection criteria.

There are 137,204 events reconstructed in the case of this reaction. The cross section for this reaction is determined as a cross check. As shown in fig. 3.9 this cross section is in good agreement with available GRAAL data [133] up to

s = 1.75 GeV. The data presented here contributes new data points above

s = 1.9 GeV. It is interesting to note that the production strength of both final states is almost the same above

s = 2.2 GeV and the maximum of the 0η cross section is about half of that for 0π0. The total cross section

a better extrapolation over the acceptance hole in the forward direction.

[MeV]

s

1600 1800 2000 2200 2400

b]µ [σ

2 4 6 8 10 12 14

Total cross sections with 20 MeV bin size

(exp) η π0

p γp

(pwa) η π0

p γp

π0

π0

p γp

(GRAAL) π0

π0

p γp

Total cross sections with 20 MeV bin size

Figure 3.9: Measured total cross sections for theγp→pπ0π0andγp→pπ0η(only statistical errors are shown)

Only statistical errors are shown in fig. 3.9. The possible sources of systematic errors are:

The background is estimated to 9% of the η peak in fig. 3.8. The background is sub-tracted; the error is estimated to 2%.

The error due to the hole in the acceptance in the forward and backward directions is estimated to be on the order of 10%.

The error in the electron energy measurement is established to be ±1%, an error on the order of 1% is estimated for the cross section.

During measurements the target center was shifted -(6.5±1.5) mm along the z-axis (beam axis). This shift was measured by two independent methods; it was measured by employing a laser technique [134] and it was investigated with the help of the vertex kinematic fit [127] and Monte Carlo simulations [114, 120]. The uncertainties in the position of the target result in the systematic error on the order of 1-2%.

Changes in the vertical position of the beam can lead to changes of the wire to energy relation; changes in the horizontal position lead toφasymmetries in the Crystal Barrel.

The impact of these asymmetries was investigated [114, 120]. The systematic error is on the order of 1%.

Calorimeter calibration slightly depends on the target position. The corresponding error however can be neglected.

The material between the target and calorimeter was implemented in the MC and carefully investigated [114, 120]. The error in the definition of the light isolating foil (for the inner detector) thickness is considered to be±1 mm and contributes systematic error for the total cross section on the order of 1-2%.

These systematic errors are quadratically added including a reconstruction error estimated to 5%. The total error thus ranges from 12-15% depending on energy. The systematic error due to flux normalization is estimated to be about 15% [114, 120] for this data set giving the main contribution to the systematic error. Flux error quadratically added with other systematic errors results in an error range of 18-21%.