X-ray microscopy with 100nm resolution and 3D imaging of the objects internal microstructure by microCT attachment for SEM.

A. Sasov, B. Pauwels, P. Bruyndonckx, and X. Liu SkyScan, Kartuizersweg 3B, 2550 Kontich, Belgium

sasov@skyscan.be

Keywords: SEM, X-ray microscopy, microtomography, micro-CT, 3D reconstruction

A standard SEM image displays the surface of an object as a two-dimensional flat picture. An inexpensive microCT attachment for any SEM has been developed to add a new imaging modality for X-ray microscopy microtomography - visualization of the true 3D microstructure inside the object without any physical cut or sample preparation [1,2].

Fig.1 shows a block-diagram of the microCT attachment for SEM. The objective lens of the SEM (1) focuses the electron beam (2) on the surface of a metal target (3). The target produces X-ray radiation (4) which passes through the object (5) and is collected by an X-ray camera (6). To acquire different angular projections the sample is mounted on a the precision rotation stage. X-ray magnification can be adjusted by changing the target-sample distance using an integrated motorized linear stage. All parts of the system except of X-ray camera are combined in one microscanner, which can be easily installed in the SEM instead of standard object stage. The X-ray camera uses a deeply cooled CCD for direct X-ray detection. The microCT attachment doesn't require any connection to the SEM electronics and works completely independently.

Using a bulk metal target allows X-ray imaging and micro-CT reconstruction with a spatial resolution down to 0.7-1.2 microns (dependent on the target material). In this case the spatial resolution is limited by the X-ray generation volume inside the target. Switching to a thin layer target (200nm Au) allows improving the spatial resolution down to 100nm. Fig.2.

shows an X-ray image of a star test pattern with a clear modulation at the 100nm level.

During scanning a computer controls the object rotation and acquires a number of X- ray angular shadow projections of the object's internal microstructure. The reconstruction program back-projects in the computer memory all acquired images and creates a set of virtual slices through the object. The results can be displayed as a slice-by slice movie or rendered to a realistic 3D model. A specially developed software allows calculating numerical parameters of internal morphology in 2D and 3D. For micro-CT reconstruction the spatial resolution is limited not only by the spot size of the X-ray emission point, but also by the accuracy in object rotation. Our reconstruction software includes an iterative alignment procedure for compensation of mechanical inaccuracies in the object rotation.

Fig.3 shows an application of the microCT attachment for SEM: non-invasive reconstruction of the internal structure of the leg of a mosquito. There are three orthogonal virtual slices with 490nm pixel size.

The developed microCT attachment for SEM opens new possibility for non- destructive 3D imaging in wide range of applications using any conventional SEM.

1. J. Cazaux et al. Journal de Physique IV 03, C7 (1993) C7-2099-C7-2104

2. A Sasov, Journal of Microscopy, vol.147, pt.2 (1987), pp.169-192.

I3.P118 - 93 - MC2009

G. Kothleitner, M. Leisch (Eds.): MC2009, Vol. 1: Instrumentation and Methodology, DOI: 10.3217/978-3-85125-062-6-045, © Verlag der TU Graz 2009

(2)

Figure 1. Block-diagram of the microCT attachment for SEM :

1 - SEM objective lens 2 - electron beam 3 - metal target 4 - X-ray beam 5 - object

6 - X-ray camera

Figure 2. X-ray image of the test pattern (NTT- AT type ATN/XRESO-50HC, Japan). Central ring contains pattern with 50nm features (cannot be resolved) and following rings with 100nm, 200nm, 500nm strips correspondingly.

Figure 3.

MicroCT reconstruction of the internal microstructure of the leg of a mosquito: three orthogonal virtual slices with 490nm pixel size.

MC2009 - 94 - I3.P118

G. Kothleitner, M. Leisch (Eds.): MC2009, Vol. 1: Instrumentation and Methodology, DOI: 10.3217/978-3-85125-062-6-045, © Verlag der TU Graz 2009

Referenzen

ÄHNLICHE DOKUMENTE

The most widely used X-ray optical elements when it comes to high resolution X-ray microscopy, especially in the soft X-ray regime, are diffractive optics in the form of Fresnel

Diffraction and coherence In X-ray diffraction imaging fringes or diffraction patterns can be detected due to the interference on an ordered system. If the X-rays are highly

A Skyscan 1172 MicroCT system was used to assess the internal and external morphology of the Australian stingless bee Trigona carbonaria with particular focus on the

In this study, we aim at (1) developing and test- ing new approaches in 2D X-ray scanning technology (i.e., increasing image resolution beyond the industrial standard of 25  µm

In this direction, it is necessary to integrate three fields of research, the photoresists production (polymer science), the X-ray lithography technology and the X-ray

The moderate requirements on coherence and monochromaticity, the possibility to make large gratings of high quality and efficiency, and the easy alignment and low sensitivity of

Likewise, the DPC image shown in Figure 2f reveals complementary details of the soft tissue structure surrounding the otoliths, whereas only the highly absorbing structures

To solve for a given field of view from a single diffraction pattern, the detector must have a pixel size (in the Fraunhofer plane) which is inversely proportional to the size of