Need to clean a sensor….

Or it got really cold.. what now?

By Alfred Wegener Institute

Sandra Tippenhauer, Gereon Budeus, Andreas Wisotzki, Gerd Rohardt Last changes 19 09 2019

In this document we describe how you should clean the CTD and its sensors. Comments and feedback for improving this document are very welcome.

At normal temperatures, you should flush the CTD sensors using clean ocean water. The water for flushing should be collected from deep water sample leftovers from other stations. If possible, filter the water using a 0.5 micron filter. If you do not have deep water you can use clean Oceanwater from OSIL (found in AWI013). Attach the syringes and make sure that the water stays in the sensors.

On Polarstern it should be fine to keep the CTD above freezing temperatures in between casts. Frequently check whether the doors and the curtains are closed. If it gets too cold in the Abfüllraum you need to dry the sensors. This applies also to the sensors in Ocean City. If you cannot make sure that the sensors will not freeze, dry them.

To dry the sensors, flush them with Bio-Ethanol. There is a 5L bottle of Bio-Ethanol in the dangerous goods store. It has the label CTD10.

There is a 1L bottle in AWI013 to get smaller amounts of the Bio-Ethanol.

Flush the sensors with Bio-Ethanol only once. That is enough. Wait until all Bio-Ethanol is out of the sensor again and dry the sensor using an aquarium pump (see picture). You find this pump in AWI013. Get a gentle stream of air though the sensor until it is dry.

Before the next measurement

If you dried sensors, you need to prepare them before you can measure again. Make sure that it is warm enough to prevent freezing (especially in ocean city).

Fill the syringes with clean ocean water and push it into the sensors. Make sure that the water stays inside the conductivity cell. Leave the water in there for at least 30 minutes before you take the first profile. If the data look strange, increase the time frame.

Oil or bio-fouling

If there is biology growing in the sensors, or the sensor is contaminated by oil, clean them with Triton X. Check the Seabird application note clean_appnote2DMar14_0.pdf for details. On the ship we have clean water (Milli-Q) and Triton X available, which was sufficient for cleaning in the past.

Referenzen

ÄHNLICHE DOKUMENTE

Welsh Water (Dŵr Cymru), Severn Trent Water, South West Water, Yorkshire Water and Anglian Water utilised paid social media promotion or geo-targeted posting to ensure they

We focus on these two water tech- nologies, because they are fast-growing technologies suiting both water quantity and quality demands (Elimelech and Phillip 2011, Gude 2017)

The application of the water planetary boundary at these subglobal contexts occurs via two approaches: (i) calculating fair shares, in which local water cycle modi fi cations

policies to water availability, and the cost, energy and emissions impacts of future infrastructure

Expansion pathways for advanced water treatment (i.e., wastewater recycling and desalination) are incorporated into the water sector transformations to supply increasing future

sustainability indicator will also depend on the ability to model the feed-backs from changes in the rest of the system onto

“to ensure the transition to democracy, including through promoting, facilitating and providing technical advice and assistance to a single, inclusive and transparent

37 year-old Maude Mondesir, Janen’s water committee treasurer, can also see changes: „The availability of services sup- plying water and sanitation are starting to