PEEM load-lock Manual

07.06.2005

I:\manuals\Load_LockManual2.doc - 1/2 -

1.) Load sample into Prepchamber

How to load a sample using the ELMITEC load lock for transfer into the PEEM.

Location

• The load lock is located below the prepchamber on the left-hand side of the PEEM.

• The forevacuum buffer is in the rack WSLA 13.0.55

1.1.) Venting the load-lock

Venting the load-lock to insert the new sample.

o Turn N2 gas ON (behind hutch) and check for flow.

o Set Buffer controller to P2 = ACTIVE.

o Check that valve separating load-lock / prepchamber and prepchamber / mainchamber are CLOSED.

o Turn TMP controller (DCU) Power = OFF o Wait until load-lock is vented.

Vent valve opens once speed is below 1350/min, after ~3 min.

Buffer controller

TMP power Buffer1

pressure

Buffer2 pressure

Load- lock

2 Screws

(2)

PEEM load-lock Manual

07.06.2005

I:\manuals\Load_LockManual2.doc - 2/2 - o Open CF40 flange

o Turn N2 gas OFF (behind hutch).

1.2.) Install new sample

Install new sample and prepare for pump-down.

o Close CF40 flange (two M6 screws and VITON gasket).

o Set Buffer controller to P1 + P2 = ACTIVE.

Wait 50sec until valve opens (Vent 1 lights up).

o Turn TMP controller (DCU) Power = ON

Check that pressure goes down (10-2 mbar after ~ 1min).

o When PLOAD < 2 10-6mbar transfer sample.

Known problems:

¾ Sometimes the vent valve of the TMP does not close. Turn TMP controller OFF / ON

Referenzen

ÄHNLICHE DOKUMENTE

Im ES4408.1 Load Chassis können insgesamt sieben Boards vom Typ ES4434.1 Configurable Load Board oder ES4435.1 Current Source Load Board eingesetzt werden.. Das ES4434.1

Viral load testing and CD4 counts CD4 testing has a role to play in the context of diagnosing opportunistic infections and managing some aspects of treatment; however, people living

What is needed to perform the calculations for the future load profiles is a database that shows for every single building in the city, how many households are in the building,

ADVANCED SEDIMENT MANAGEMENT IN MOUNTAINOUS BASIN In Jo-Gan-ji River, the monitoring tools and the systems for data collection for evaluating sediment runoff were prepared

Through monitoring of those events, it was found that debris flow occurred on June 15th had several surges during a flood and temporal changes of coarse component

In Section 3, we present a high performance distributed computing approach for molecular fragment mining and the adopted dynamic load balancing policy.. Section 5 describes

Sector 17: The communication area is used to pass device informa- tion (device address, Series/l storage address) from the diskette read program to the attachment load or

A tachometer as defined in claim 8, wherein: the means for generating a voltage includes a capacitor connected between two fixed voltage levels, the capacitor being con