Noise Analysis of Amplitude and Phase

reconstructed from High Signal Resolution Electron Holograms

Falk Röder1, Axel Lubk1, Dorin Geiger1 and Hannes Lichte1

1. Technische Universität Dresden, Institut für Strukturphysik, Triebenberg Labor, 01062 Dresden, Germany

Falk.Roeder@Triebenberg.de

Keywords: electron holography, standard deviation, noise transfer, averaging

In a TEM, the wave and in particular the phase is modulated very sensitively and by electromagnetic fields. Hence the phase reconstructed by means of Electron Holography provides a powerful tool to investigate modulations of electric or magnetic fields in or around solids [1], (Figure 2a). The reconstructed amplitude damped by e.g. inelastic scattering events additionally allows an estimation of the local thickness. This is useful for extracting mean potentials from the phase [2].

Nevertheless the ability to obtain information from amplitude and phase strongly depends on the according signal-to-noise ratios. F. Lenz [3] and A. Harscher, H. Lichte [4]

derived analytical expression for the standard deviations of the two mentioned quantities.

However, the considerable influence of the scintillator in the detector on the noise was neglected. Hence a more realistic noise model requires the incorporation of the correlation of statistical properties between different pixels in the hologram due to the scintillator before transferring to the three reconstructed channels, i.e. center band, amplitude and phase.

We investigate the noise properties experimentally by evaluating the statistics of amplitude and phase images from a series of N holograms in the same area of magnetic particles. That provides direct access to the local noise in terms of the experimental standard deviation in each reconstructed pixel (Figures 1b and 2b). The thereby obtained standard deviation of the amplitude in the particle is lower than in the vacuum region (Figure 2b), corresponding qualitatively to the prediction of Harscher and Lichte [4]; the analytic behavior however cannot be confirmed in this experiment, due to neglecting the noise correlation by the scintilator. A model including the amplifying and broadening influences of the detector on the noise (i.e. incorporating Noise Transfer Function) will be represented in this contribution.

We will discuss the advantage that averaging over all N reconstructed amplitude and phase images at the same object position reduces the amount of noise for both by a factor of 1/sqrt(N) (Figures 1a and 2a) and provides automatic access to the standard deviation of the mean values, which can be employed as error bars. This is helpful to check the reliability of the reconstructed signals in all applications of holography.

1. H. Lichte, P. Formanek, A. Lenk, M. Linck, C. Matzeck, M. Lehmann, P. Simon:

Annual Review of Materials Research, Vol. 37 (2007), 539-588

2. M. R. McCartney, M. Gajdardziska-Josifovska: Ultramicroscopy 53 (1994) 283-289 3. F. Lenz: Statistics of phase and contrast determination in electron holograms,

Optik 79 No.1 13-14, 1988

4. A. Harscher, H. Lichte: Experimental study of amplitude and phase detection limit in electron holography, Ultramicroscopy 64 (1996) 57-66

I1.116 - 19 - MC2009

G. Kothleitner, M. Leisch (Eds.): MC2009, Vol. 1: Instrumentation and Methodology, DOI: 10.3217/978-3-85125-062-6-008, © Verlag der TU Graz 2009

(2)

Figure 1a. Mean amplitude of BaFe12O19 particleswith according profile

Figure 1b. Standard deviation of mean amplitude of BaFe12O19 particles with according profile indicating a decreasing for low amplitude values

Figure 2a. Mean phase image of BaFe12O19 particleswith according profile showing phase shifts due to magnetic fields in an around the particles.

Figure 2b. Standard deviation of mean phase image of BaFe12O19 particleswith according profile revealing an increase of phase noise in the particle.

Α

σ

Α

φ

σ

φ

MC2009 - 20 - I1.116

G. Kothleitner, M. Leisch (Eds.): MC2009, Vol. 1: Instrumentation and Methodology, DOI: 10.3217/978-3-85125-062-6-008, © Verlag der TU Graz 2009

Referenzen

ÄHNLICHE DOKUMENTE

In this thesis, we describe our scientific research and discoveries that shed light on quan- tum effects in the phase and amplitude dynamics of quantum self-oscillators, and

lectures where students are taught the content and methodol- ogy of their future profession are accompanied by two practical training periods of several weeks (6 weeks and 4

At temperatures below the transition temperature diffraction experiments on single- crystals of TiI 3 evidenced the occurrence of superstructure reflections whose inten- sities were

where W 60 is the experimental weight (g) of the plastic pieces recovered after 60 days of composting, 10 g is the ini- tial weight of the plastic sample, water (%) and

Key words: Terbium(III) Bromide; Potassium Bromide; Enthalpy; Phase Diagram; Differential Scanning Calorimetry; Eutectic; Compound;

In this paper the properties of the high frequency contribution of the dielectric relaxation (up to 3 GHz) in the vicinity of the 168 and 134 K phase transitions of MABA crystals

Zebra finches are one of the few songbird species in which variation in song amplitude between males has been studied (Brumm and Slater 2006b; Brumm 2009). During

Influence of amplitude on the period at the propagation depth H p as measured in Lake Constance for 2 solitary wave trains consisting of several solitary waves, which occurred