Treatment Planning for Pituitary Adenomas –

A Comparison of Flatness Filter Free and Flat Beam Plans

Kontakt E-Mail: marius.treutwein@ukr.de

DEGRO 24. Jahrestagung (21.-24.Juni 2018 in Leipzig)

Archived on the repository of the University of Regensburg at: http://epub.uni-regensburg.de/37067/

Introduction

Radiotherapy of pituitary adenomas is often applied as postoperative therapy of tumors that cannot be removed completely. Fluence

modulating techniques like IMRT or VMAT support the use of the

flatness filter free (FFF) mode in modern linear accelerators. In this mode a considerably higher dose rate is achieved by omitting the flatness filter. This planning study compares different plans for

patients with pituitary adenoma. The plans were optimized using both modes: flattened beam (FB) and FFF. The aim of this study is to

evaluate the plan quality.

Material and methods

Data sets of 11 patients with pituitary adenoma were used for this

retrospective planning study. The objectives for the PTV were set to a minimum dose of 49.4Gy and a maximum dose of 51.4Gy in 28

fractions, aiming for a fraction dose of 1.8Gy. Further objectives were set for the following regions of interest (ROI): brain, brainstem,

chiasm, and both lenses, bulbs, lacrimal glands, and parotids (table 1). The planning was performed with the treatment planning system (TPS) Oncentra External Beam v4.5 (Elekta AB, Sweden) using the

collapsed cone algorithm. In both modes (FB and FFF) IMRT plans with nine equispaced coplanar fields were generated (IMRT9); in a second variant a tenth non-coplanar field was added (IMRT10).

Similarly two VMAT plans were optimized: one single arc rotation

(182°-178°)(VMAT1), and the second with an added half rotation in the sagittal patient plane (0°-180°)(VMAT2). The applied linac

Synergy Agility (Elekta AB, Sweden) offers a dose rate of 550MU/min (FB) and 1700MU/min (FFF). The leaves have a width of 5mm

projected to the isocenter.

Table 1. Treatment Planning Objectives

The following parameters were evaluated: the average dose D Av to the PTV, homogeneity index HI, and the conformity index CI. For all plans the observance of the objectives for the organs at risk given in table 1 was investigated. The data were analyzed with SPSS® (IBM, USA) to

generate diagrams and visualize statistical parameters.

Results

For both modes FB and FFF the dose volume statistics are very close.

With some minor exceptions the objectives for the ROIs are met. Only the maximum dose to the chiasm is slightly exceeded in most cases, as it is part of the PTV. Regarding the PTV (HI and CI) pronounced

differences are found in the comparison of IMRT and VMAT (diagram 1 and 2) showing a benefit for VMAT. The distinction between

coplanar and non-coplanar techniques is less important.

Diagram 1. Boxplot of the conformity index CI of the different plan variants, showing

mean value, inner quartile (box) and outer quartile (bar), and a few outliers (circles) and extreme value (asterisk). FB yellow, FFF blue.

Diagram 1. Boxplot of the homogeneity index HI of the different plan variants, showing mean value, inner quartile (box) and outer quartile (bar). FB yellow, FFF blue.

Discussion

The difference in the plan quality of FB and FFF plans is of no clinical importance. A decision for the preferred treatment technique will

therefore be taken based on measurements of the delivery time and peripheral dose which will follow. As the plan quality of VMAT is

superior to IMRT, the decision will be made between VMAT FB and VMAT FFF. All non-coplanar techniques will require more linac time due to couch rotations without showing advantages in plan quality.

Acknowledgement

This work was supported by the Bavarian State Ministry of the Environment and Consumer Protection (Bayerisches Ministerium für Umwelt und

Verbraucherschutz)

Marius Treutwein 1 , Felix Steger 1 , Oliver Kölbl 1 , Barbara Dobler 1

1

Regensburg University Medical Center, Radiotherapy Department, Regensburg, Germany

Organ at Risk max Dose in Gy max Volume in %

Brain 20 20

30 10

40 5

Brainstem 51.4 0

Chiasm 50 0

Lens 15 0

Bulb 35 50

Lacrimal gland 20 50

Parotid 30 50

Referenzen

ÄHNLICHE DOKUMENTE

meteorite contains C-rich material at some individual measurement po- sitions. The other points can be identi fi ed as a mixture of melilite and albite. 6d) show that the

In 1988, Polish literary scholar Mirosława Czarnecka described the status of female wri- ters within world literature as follows: »The woman as author appears in the tradition of

An increase in relative income poverty might hint at the distributional effects of involuntary unemployment, since it can be supposed to mainly concern households supplying

Sie haben allerdings wesentlich mehr Aussagekraft über die Verteilung der Daten, da sie diese sozusagen in vier gleich große „Häppchen“ aufteilen: Der Median teilt die Daten in

To directly examine the magnitude of errors that can occur, we use dual whole-cell current- and voltage-clamp recordings from the soma of the pair of easily accessible,

LH gonadotroph pituitary adenoma; PITAD ACTH: adrenocortico- troph hormone producing pituitary adenoma, PITAD PRL prolacti- noma; PITAD TSH thyrotroph hormone producing

As stated in the previous section, algorithms basically differ in three ar- eas: by the number of features used (intra-pair versus inter- pair and intersections only

In particular, marketing planning frequently lacks formal linkages to corporate planning, it is narrow in its generation and evaluation of strategies, it lacks formal schemes